A comparative investigation of the quality of radiographs produced by portable handheld and fixed x-ray units

To evaluate the impact of handheld portable intraoral diagnostic X-ray equipment on radiographic quality and to compare their radiographs with radiographs obtained using fixed intraoral X-ray equipment. For brightness, noise, and uniformity, radiographs of an acrylic block were obtained with two complementary metal oxide semiconductor (CMOS) sensors (Digora Toto and Snapshot). Six radiographs for each sensor were obtained using the Eagle handheld portable X-ray equipment set at 60 kVp, 2.5 mA, and 0.45 s. Then, six radiographs were obtained with the Focus fixed X-ray equipment set at 60 kVp, 7 mA, and 0.16 s. Mean and standard deviation of the gray values were evaluated. For contrast, radiographs of an aluminum step-wedge were obtained using both sensors, X-ray equipment, and the aforementioned acquisition parameters. The percentage of contrast variation was evaluated. The comparison between the equipment was evaluated for its influence on radiographic quality using Student's t-test (p < 0.05). The handheld portable X-ray equipment produced radiographs with higher brightness and lower contrast regardless of the CMOS sensor used (p<0.0001). Noise levels were higher with the handheld portable X-ray equipment than with the fixed X-ray equipment using the Digora Toto system, and lower with the Snapshot system (p < 0.05). No significant differences were found for uniformity. Therefore, radiographs obtained with a handheld portable intraoral diagnostic X-ray equipment have higher brightness and lower contrast compared to those obtained with fixed X-ray equipment. While the effect on noise levels varies depending on the digital system used, the uniformity remains unaffected by the type of X-ray equipment.

Authors

Institutions

Publication Details

Journal
Brazilian Dental Journal
Published
2025-01-01
DOI
https://doi.org/10.1590/0103-644020256319
Primary Topic
Dental Radiography and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A comparative investigation of the quality of radiographs produced by portable handheld and fixed x-ray units

Rocharles Cavalcante Fontenele, Deborah Queiroz Freitas, Débora Costa Ruiz, Matheus L. Oliveira et al.
Brazilian Dental Journal
Dental Radiography and Imaging
article

A comparative investigation of the quality of radiographs produced by portable handheld and fixed x-ray units

Rocharles Cavalcante Fontenele, Deborah Queiroz Freitas, Débora Costa Ruiz, Matheus L. Oliveira, Francisco Haiter‐Neto
article en

Abstract

To evaluate the impact of handheld portable intraoral diagnostic X-ray equipment on radiographic quality and to compare their radiographs with radiographs obtained using fixed intraoral X-ray equipment. For brightness, noise, and uniformity, radiographs of an acrylic block were obtained with two complementary metal oxide semiconductor (CMOS) sensors (Digora Toto and Snapshot). Six radiographs for each sensor were obtained using the Eagle handheld portable X-ray equipment set at 60 kVp, 2.5 mA, and 0.45 s. Then, six radiographs were obtained with the Focus fixed X-ray equipment set at 60 kVp, 7 mA, and 0.16 s. Mean and standard deviation of the gray values were evaluated. For contrast, radiographs of an aluminum step-wedge were obtained using both sensors, X-ray equipment, and the aforementioned acquisition parameters. The percentage of contrast variation was evaluated. The comparison between the equipment was evaluated for its influence on radiographic quality using Student's t-test (p < 0.05). The handheld portable X-ray equipment produced radiographs with higher brightness and lower contrast regardless of the CMOS sensor used (p<0.0001). Noise levels were higher with the handheld portable X-ray equipment than with the fixed X-ray equipment using the Digora Toto system, and lower with the Snapshot system (p < 0.05). No significant differences were found for uniformity. Therefore, radiographs obtained with a handheld portable intraoral diagnostic X-ray equipment have higher brightness and lower contrast compared to those obtained with fixed X-ray equipment. While the effect on noise levels varies depending on the digital system used, the uniformity remains unaffected by the type of X-ray equipment.

Brazilian Dental JournalVol. 36
Universidade Estadual de Campinas (UNICAMP) (BR)
Openalex Percentile: Top 58%
Dental Radiography and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.