Evaluation of the Effect of Decreased Surrogate Concentrations on the Calculation Value of Recovery Rates Using Multiple Regression Analysis and Calibration Curves on LC/MS Analysis

The effect of varying the surrogate concentration on the calculation value of the recovery rate was investigated by LC/MS and data analysis methods using <i>o</i>-toluidine as a model compound. The low-concentration mixture samples, which simulated the decrease in concentrations of analyte (<i>o</i>-toluidine, <i>m</i>/<i>z</i> 108) and surrogate (<i>o</i>-toluidine-<i>d</i> <sub>3</sub>, <i>m</i>/<i>z</i> 111), were prepared by diluting the standard mixture. Furthermore, the relationship between the prepared and detected analyte/surrogate concentrations in these model samples was examined, both in the presence and absence of the internal standard (IS, <i>o</i>-toluidine-4,6-<i>d</i> <sub>2</sub>, <i>m</i>/<i>z</i> 110). Although the same proportion of components was present at each concentration level, the changes in the analyte/surrogate peak area proportions were more significant at lower sample concentrations. Moreover, this effect was more pronounced in samples containing both IS and surrogate than in those containing only the surrogate, with the proportion decreasing to approximately 60% owing to the <i>m</i>/<i>z</i> 111 peak from the IS. Among the methods for calculating the recovery rate, multiple regression analysis using analyte/IS and surrogate/IS could calculate the recovery rate at low concentrations, whereas the method using two calibration curves with analyte/IS and surrogate/IS performed the same or slightly worse. Furthermore, upon applying the methods to solid-phase extraction samples, the results were the same. The accuracy of the recovery rate at each concentration level was evaluated by checking whether the detected (measured) analyte/surrogate ratio matched the prepared (theoretical) ratio in the standard mixture.

Authors

Institutions

Publication Details

Journal
ACS Omega
Published
2025-10-24
DOI
https://doi.org/10.1021/acsomega.5c08745
Primary Topic
Analytical chemistry methods development
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of the Effect of Decreased Surrogate Concentrations on the Calculation Value of Recovery Rates Using Multiple Regression Analysis and Calibration Curves on LC/MS Analysis

Naoko Inoue
ACS Omega
Analytical chemistry methods development
article

Evaluation of the Effect of Decreased Surrogate Concentrations on the Calculation Value of Recovery Rates Using Multiple Regression Analysis and Calibration Curves on LC/MS Analysis

Naoko Inoue
article en

Abstract

The effect of varying the surrogate concentration on the calculation value of the recovery rate was investigated by LC/MS and data analysis methods using <i>o</i>-toluidine as a model compound. The low-concentration mixture samples, which simulated the decrease in concentrations of analyte (<i>o</i>-toluidine, <i>m</i>/<i>z</i> 108) and surrogate (<i>o</i>-toluidine-<i>d</i> <sub>3</sub>, <i>m</i>/<i>z</i> 111), were prepared by diluting the standard mixture. Furthermore, the relationship between the prepared and detected analyte/surrogate concentrations in these model samples was examined, both in the presence and absence of the internal standard (IS, <i>o</i>-toluidine-4,6-<i>d</i> <sub>2</sub>, <i>m</i>/<i>z</i> 110). Although the same proportion of components was present at each concentration level, the changes in the analyte/surrogate peak area proportions were more significant at lower sample concentrations. Moreover, this effect was more pronounced in samples containing both IS and surrogate than in those containing only the surrogate, with the proportion decreasing to approximately 60% owing to the <i>m</i>/<i>z</i> 111 peak from the IS. Among the methods for calculating the recovery rate, multiple regression analysis using analyte/IS and surrogate/IS could calculate the recovery rate at low concentrations, whereas the method using two calibration curves with analyte/IS and surrogate/IS performed the same or slightly worse. Furthermore, upon applying the methods to solid-phase extraction samples, the results were the same. The accuracy of the recovery rate at each concentration level was evaluated by checking whether the detected (measured) analyte/surrogate ratio matched the prepared (theoretical) ratio in the standard mixture.

ACS OmegaVol. 10(43)
National Institute of Occupational Safety and Health (JP)
Openalex Percentile: Top 65%
Analytical chemistry methods development
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.