COMPUTATIONAL PREDICTION OF THE IMPACT OF SCAPULAR NOTCHING ON ARTICULAR WEAR RATE

Wear of the intended articular surface and cup damage due to scapular impingement are known issues in reverse total shoulder arthroplasty (RTSA), but the relationship between them has not yet been fully investigated. This study includes the development of a computational wear model for the prediction of RTSA articular wear rates and implementation of this model to examine the interaction between cup damage due to scapular notching and articular wear. Three-dimensional models of 38mm diameter, 8.75mm cup depth, RTSA implants were constructed assuming 0.08mm radial clearance, and scapular notching damage was simulated as a circular arc with a width of 14mm and a depth of 3mm on the inferomedial edge of the cup. Finite element methods were used to ascertain the contact pressure across the articular surface of the humeral cup representing the relative motion and loading applied during previously performed in-vitro wear testing, which included a circumduction motion with an applied load ranging from 814–913N. Contact pressure was then input into a computational wear model which incorporated both the Archard wear model and the pressure-independent polyethylene (PE) wear model proposed by Liu et al. [1] to estimate the articular wear rate. The wear depth per million cycles (MC) across the surface of the cup was then integrated to determine the volumetric wear rate for both the unnotched cup and the cup with simulated scapular notching. The computational model predicted a slightly lower volumetric wear rate in the notched cup compared to the intact cup: a decrease of four percent using the Archard wear model (34.0mm3/MC unnotched vs. 32.6mm3/MC notched) and nine percent using the PE wear model (62.3mm3/MC unnotched vs. 56.6mm3/MC notched). This is consistent with the in-vitro results shown by Griffiths et al. who found a decrease of 26% (31.2mm3/MC unnotched vs. 23.0mm3/MC notched [2]), and Langohr et al. who found a decrease of eight percent (42.0mm3/MC unnotched vs. 38.8mm3/MC notched [3]). The similarity between the trends predicted by the computational model and the results of in-vitro wear testing demonstrates that the computational model is a valuable tool to investigate wear in RTSA implants. Its comparatively low cost and short time to yield results make it useful for evaluating changes in implant design and testing protocols before proceeding to full-scale in-vitro testing using only the most promising selections from the in-silico testing.

Authors

Institutions

Publication Details

Journal
Orthopaedic Proceedings
Published
2025-10-22
DOI
https://doi.org/10.1302/1358-992x.2025.10.016
Primary Topic
Muscle activation and electromyography studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

COMPUTATIONAL PREDICTION OF THE IMPACT OF SCAPULAR NOTCHING ON ARTICULAR WEAR RATE

G. Daniel G. Langohr, Emma Badowski
Orthopaedic Proceedings
Muscle activation and electromyography studies
article

COMPUTATIONAL PREDICTION OF THE IMPACT OF SCAPULAR NOTCHING ON ARTICULAR WEAR RATE

G. Daniel G. Langohr, Emma Badowski
article en

Abstract

Wear of the intended articular surface and cup damage due to scapular impingement are known issues in reverse total shoulder arthroplasty (RTSA), but the relationship between them has not yet been fully investigated. This study includes the development of a computational wear model for the prediction of RTSA articular wear rates and implementation of this model to examine the interaction between cup damage due to scapular notching and articular wear. Three-dimensional models of 38mm diameter, 8.75mm cup depth, RTSA implants were constructed assuming 0.08mm radial clearance, and scapular notching damage was simulated as a circular arc with a width of 14mm and a depth of 3mm on the inferomedial edge of the cup. Finite element methods were used to ascertain the contact pressure across the articular surface of the humeral cup representing the relative motion and loading applied during previously performed in-vitro wear testing, which included a circumduction motion with an applied load ranging from 814–913N. Contact pressure was then input into a computational wear model which incorporated both the Archard wear model and the pressure-independent polyethylene (PE) wear model proposed by Liu et al. [1] to estimate the articular wear rate. The wear depth per million cycles (MC) across the surface of the cup was then integrated to determine the volumetric wear rate for both the unnotched cup and the cup with simulated scapular notching. The computational model predicted a slightly lower volumetric wear rate in the notched cup compared to the intact cup: a decrease of four percent using the Archard wear model (34.0mm3/MC unnotched vs. 32.6mm3/MC notched) and nine percent using the PE wear model (62.3mm3/MC unnotched vs. 56.6mm3/MC notched). This is consistent with the in-vitro results shown by Griffiths et al. who found a decrease of 26% (31.2mm3/MC unnotched vs. 23.0mm3/MC notched [2]), and Langohr et al. who found a decrease of eight percent (42.0mm3/MC unnotched vs. 38.8mm3/MC notched [3]). The similarity between the trends predicted by the computational model and the results of in-vitro wear testing demonstrates that the computational model is a valuable tool to investigate wear in RTSA implants. Its comparatively low cost and short time to yield results make it useful for evaluating changes in implant design and testing protocols before proceeding to full-scale in-vitro testing using only the most promising selections from the in-silico testing.

Orthopaedic ProceedingsVol. 107-B(SUPP_10)
Western University (CA)
Openalex Percentile: Top 70%
Muscle activation and electromyography studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.