Fundamental Sensor Response Time Limitations of Practical Air Temperature Measurement
Abstract Air temperature measurements in naturally ventilated thermometer screens underpin the instrumental climate record. Increasing automation is, however, revealing limitations. One is through thermometer time response, especially in light winds or calm conditions, often at the daily temperature minimum. The exponential time response for thermometers enclosed within a Stevenson screen is a key parameter, but poorly known. Here, is evaluated in a practical experimental situation against the World Meteorological Organization (WMO)'s recommended sensor s. We find increases with sensor diameter , with only a = 2 mm sensor meeting WMO expectations, even then requiring ambient wind speeds . Typical = 4 mm sensors never meet the criterion when either force‐ or naturally ventilated, with mins in a naturally ventilated arrangement under calm conditions. Inadequate will lead to underestimation of the diurnal temperature range or other local measures derived from daily temperature maxima and minima.