Comparison of Pulse Oximetry and Blood Pressure Measurements Between Sensor Locations in Children During Dental Procedures.

<b>Purpose:</b> To compare oxygen saturation measurements when pulse oximeter sensors were simultaneously placed on the index finger and the second toe, and compare blood pressure recordings with non-invasive blood pressure (NIBP) cuffs simultaneously placed on the upper arm and the lower calf in children receiving general anesthesia. <b>Methods:</b> A total of 50 children, aged three to nine years, defined as ASA Class I or II receiving comprehensive dental treatment in the operating room with general anesthesia, were included. Pulse oximeter sensors were simultaneously placed on each subject's index finger and ipsilateral second toe. Non-invasive blood pressure cuffs were simultaneously placed on the upper arm and ipsilateral lower calf above the ankle. The same anesthesia regimen was used on all patients. Following induction of general anesthesia, measurements were simultaneously recorded every five minutes during the dental procedure. <b>Results:</b> Index finger oxygen saturation measurements were significantly lower than the second toe (P<0.001). Arm systolic blood pressure measurements were significantly lower than calf measurements (P<0.001). Arm diastolic blood pressure measurements were significantly lower than calf measurements (P<0.001). <b>Conclusions:</b> Discrepancy of vital sign measurements may be notable depending on the site from which measurements are obtained. If blood pressure is taken on the calf of a child in the Trendelenburg position, the reading will likely produce a higher systolic and diastolic measurement and should be interpreted with caution. A pulse oximeter placed on the first finger or second toe of healthy children can be regarded as interchangeable in a clinical setting.

Authors

Institutions

Publication Details

Journal
PubMed
Published
2025-09-15
Primary Topic
Non-Invasive Vital Sign Monitoring
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparison of Pulse Oximetry and Blood Pressure Measurements Between Sensor Locations in Children During Dental Procedures.

George J. Eckert, Juan F. Yepes, LaQuia A. Vinson, Allison C Scully et al.
PubMed
Non-Invasive Vital Sign Monitoring
article

Comparison of Pulse Oximetry and Blood Pressure Measurements Between Sensor Locations in Children During Dental Procedures.

George J. Eckert, Juan F. Yepes, LaQuia A. Vinson, Allison C Scully, Stephen Wilson, James E. Jones, Brianne N Cipich
article en

Abstract

<b>Purpose:</b> To compare oxygen saturation measurements when pulse oximeter sensors were simultaneously placed on the index finger and the second toe, and compare blood pressure recordings with non-invasive blood pressure (NIBP) cuffs simultaneously placed on the upper arm and the lower calf in children receiving general anesthesia. <b>Methods:</b> A total of 50 children, aged three to nine years, defined as ASA Class I or II receiving comprehensive dental treatment in the operating room with general anesthesia, were included. Pulse oximeter sensors were simultaneously placed on each subject's index finger and ipsilateral second toe. Non-invasive blood pressure cuffs were simultaneously placed on the upper arm and ipsilateral lower calf above the ankle. The same anesthesia regimen was used on all patients. Following induction of general anesthesia, measurements were simultaneously recorded every five minutes during the dental procedure. <b>Results:</b> Index finger oxygen saturation measurements were significantly lower than the second toe (P<0.001). Arm systolic blood pressure measurements were significantly lower than calf measurements (P<0.001). Arm diastolic blood pressure measurements were significantly lower than calf measurements (P<0.001). <b>Conclusions:</b> Discrepancy of vital sign measurements may be notable depending on the site from which measurements are obtained. If blood pressure is taken on the calf of a child in the Trendelenburg position, the reading will likely produce a higher systolic and diastolic measurement and should be interpreted with caution. A pulse oximeter placed on the first finger or second toe of healthy children can be regarded as interchangeable in a clinical setting.

PubMedVol. 47(5)
Nationwide Children's Hospital (US), Indiana University – Purdue University Indianapolis (US)
Openalex Percentile: Top 70%
Non-Invasive Vital Sign Monitoring
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.