First intercomparison of 241Am activity measurements using primary methods including magnetic micro-calorimetrers

Low-temperature detectors, such as Magnetic Micro-calorimeters (MMCs), exhibit several characteristics that make them strong candidates for primary activity standardization of various radionuclides. As a relatively new technique, however, many aspects of MMC-based measurements still require validation. For this purpose, <sup>241</sup>Am was selected to conduct an intercomparison of activity measurements between MMCs and other well-established conventional methods. The alpha-particle emitter <sup>241</sup>Am is well-suited to such comparisons, as it can be standardized with low uncertainty using a variety of techniques, including defined solid angle counting, coincidence counting, ionization chambers, and liquid scintillation counting (LSC). A key challenge in this context is the preparation of sources compatible with both MMCs and traditional standardization methods. An <sup>241</sup>Am source was prepared by electro-precipitation in a specially designed geometry compatible with both multi-channel MMC spectrometry and conventional methods. This source was measured by four metrology institutes - LNE-LNHB, CIEMAT, CHUV, and PTB - using different primary methods. Subsequently, the original source was divided into 21 secondary sources, which were integrated into 11 absorbers of a multi-channel MMC spectrometer for the final activity determination. Activity measurements using MMCs will be presented, along with the results of the intercomparison between different laboratories and techniques. The reported uncertainties are relatively high, in the range of a few percent, due to the rather low activity of the source and a potential loss of activity during the exercise. However, this first intercomparison provides insights and opportunities for improvement, which will be implemented in the future.

Authors

Institutions

Publication Details

Journal
Applied Radiation and Isotopes
Published
2025-10-23
DOI
https://doi.org/10.1016/j.apradiso.2025.112272
Primary Topic
Radioactive Decay and Measurement Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

First intercomparison of 241Am activity measurements using primary methods including magnetic micro-calorimetrers

F. Juget, M. Roteta, V. Peyrés, O. Nähle et al.
Applied Radiation and Isotopes
Radioactive Decay and Measurement Techniques
article

First intercomparison of 241Am activity measurements using primary methods including magnetic micro-calorimetrers

F. Juget, M. Roteta, V. Peyrés, O. Nähle, Sebastian Kempf, M. P. Takács, Sylvie Pierre, J. Beyer, Alexander Göggelmann, Valérie Lourenço, Karsten Kossert, M. Loidl, Matias Rodrigues, Michael Müller, Lucille Chambon, A.I. Sánchez-Cabezudo, Quentin Drenne, Marco Lombana
article en

Abstract

Low-temperature detectors, such as Magnetic Micro-calorimeters (MMCs), exhibit several characteristics that make them strong candidates for primary activity standardization of various radionuclides. As a relatively new technique, however, many aspects of MMC-based measurements still require validation. For this purpose, <sup>241</sup>Am was selected to conduct an intercomparison of activity measurements between MMCs and other well-established conventional methods. The alpha-particle emitter <sup>241</sup>Am is well-suited to such comparisons, as it can be standardized with low uncertainty using a variety of techniques, including defined solid angle counting, coincidence counting, ionization chambers, and liquid scintillation counting (LSC). A key challenge in this context is the preparation of sources compatible with both MMCs and traditional standardization methods. An <sup>241</sup>Am source was prepared by electro-precipitation in a specially designed geometry compatible with both multi-channel MMC spectrometry and conventional methods. This source was measured by four metrology institutes - LNE-LNHB, CIEMAT, CHUV, and PTB - using different primary methods. Subsequently, the original source was divided into 21 secondary sources, which were integrated into 11 absorbers of a multi-channel MMC spectrometer for the final activity determination. Activity measurements using MMCs will be presented, along with the results of the intercomparison between different laboratories and techniques. The reported uncertainties are relatively high, in the range of a few percent, due to the rather low activity of the source and a potential loss of activity during the exercise. However, this first intercomparison provides insights and opportunities for improvement, which will be implemented in the future.

Applied Radiation and IsotopesVol. 227
Karlsruhe Institute of Technology (DE), Physikalisch-Technische Bundesanstalt (DE), Commissariat à l'Énergie Atomique et aux Énergies Alternatives (FR), Université Paris-Saclay (FR), Laboratoire d'Intégration des Systèmes et des Technologies (FR), Laboratoire National Henri Becquerel (FR), Centro de Investigaciones Energéticas, Medioambientales y Tecnológicas (ES), Institute of Radiation Physics (CH)
Openalex Percentile: Top 60%
Radioactive Decay and Measurement Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.