TRAFFIC CAMERAS: AN EFFECTIVE AND SUSTAINABLE METHOD OF REDUCING TRAFFIC AND AIRBORNE PARTICLES DURING ARTHROPLASTY SURGERY

Prosthetic joint infections (PJI) are now the most common cause of reoperation following hip replacement surgery. Traffic in the operating room (OR) create turbulence and contaminates ultraclean air by bacterial shedding. Therefore, controlling traffic in the OR is an important strategy to prevent infection by reducing airborne particles capable of carrying bacteria. In this study, we examined (1) if the number and duration of door openings was associated with increased particle counts during arthroplasty surgery (2) if a traffic camera installed in the operating room was an effective intervention to decrease traffic and particle counts during arthroplasty surgery and (3) the effectiveness of the traffic camera over time. A prospective observational study examined all primary joint replacements at a high-volume, academic hospital between November 2021 and June 2022. Two aerosolized particle counters were used to count particles sized 0.5–10 μm, corresponding to the particles sizes that most commonly contain bacteria. One particle counting machine was positioned adjacent to the nursing back table within the sterile operative field, and the other was placed between the two OR doors that are used for personnel to enter and leave the OR during surgery. Additionally, two magnetic door counters were mounted on the two operating room doors and were used to count the number of door openings during surgery. For the intervention, we installed warning signs on the doors and traffic cameras facing each door, that took a snapshot with every door opening during surgery. A total of 50 cases were included in the study with 25 cases in the Control group and 25 cases in the Intervention group. The number of door openings/minute was 30% less in the Intervention group than in the Control group (0.7±0.2 vs 1.0±0.3, P < 0 .001). For all particle sizes, compared to the Control group, the Intervention group significantly decreased the particle counts by 26–43% in the operative field (0.5μm, p = 0.01; 0.7μm, p = 0.008; 1μm, p = 0.007; 2.5 μm, p = 0.006; 5um, p = 0.01and 10μm, p = 0.01). The particle counts between the OR doors were decreased by 2–42% in the Intervention group compared to the Control group, and the difference was significant for particles sized 0.5μm, 0.7μm, and 1μm (0.5μm, p=0.03, 0.7μm, p=0.02 and 1μm, p=0.03). A significant decrease (56–78%) was found in the average particle count in the Intervention group between the first 15 minutes and the last 15 minutes of the case compared to the Control group (0.5μm, p=0.01; 0.7μm, p=0.004; 1μm, p=0.002; 2.5μm, p=0.001; 5μm, p=0.004 and 10μm, p=0.01). The decrease in door openings and particle counts attained in the Intervention group were sustained over the entire study period. This study demonstrated that the use of traffic cameras was an effective and sustainable method to limit OR traffic during arthroplasty surgery. The use of door posters and traffic cameras significantly limited OR traffic and reduced door openings, which resulted in a sustained reduction in particle counts near the OR doors, and more importantly within the operative field.

Authors

Institutions

Publication Details

Journal
Orthopaedic Proceedings
Published
2025-10-22
DOI
https://doi.org/10.1302/1358-992x.2025.10.081
Primary Topic
Healthcare and Environmental Waste Management
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

TRAFFIC CAMERAS: AN EFFECTIVE AND SUSTAINABLE METHOD OF REDUCING TRAFFIC AND AIRBORNE PARTICLES DURING ARTHROPLASTY SURGERY

Muadh Alzeedi, Michael D. Tanzer, Adam G. Hart, Anas Nooh et al.
Orthopaedic Proceedings
Healthcare and Environmental Waste Management
article

TRAFFIC CAMERAS: AN EFFECTIVE AND SUSTAINABLE METHOD OF REDUCING TRAFFIC AND AIRBORNE PARTICLES DURING ARTHROPLASTY SURGERY

Muadh Alzeedi, Michael D. Tanzer, Adam G. Hart, Anas Nooh, Tommy Lavoie-Turcotte
article en

Abstract

Prosthetic joint infections (PJI) are now the most common cause of reoperation following hip replacement surgery. Traffic in the operating room (OR) create turbulence and contaminates ultraclean air by bacterial shedding. Therefore, controlling traffic in the OR is an important strategy to prevent infection by reducing airborne particles capable of carrying bacteria. In this study, we examined (1) if the number and duration of door openings was associated with increased particle counts during arthroplasty surgery (2) if a traffic camera installed in the operating room was an effective intervention to decrease traffic and particle counts during arthroplasty surgery and (3) the effectiveness of the traffic camera over time. A prospective observational study examined all primary joint replacements at a high-volume, academic hospital between November 2021 and June 2022. Two aerosolized particle counters were used to count particles sized 0.5–10 μm, corresponding to the particles sizes that most commonly contain bacteria. One particle counting machine was positioned adjacent to the nursing back table within the sterile operative field, and the other was placed between the two OR doors that are used for personnel to enter and leave the OR during surgery. Additionally, two magnetic door counters were mounted on the two operating room doors and were used to count the number of door openings during surgery. For the intervention, we installed warning signs on the doors and traffic cameras facing each door, that took a snapshot with every door opening during surgery. A total of 50 cases were included in the study with 25 cases in the Control group and 25 cases in the Intervention group. The number of door openings/minute was 30% less in the Intervention group than in the Control group (0.7±0.2 vs 1.0±0.3, P < 0 .001). For all particle sizes, compared to the Control group, the Intervention group significantly decreased the particle counts by 26–43% in the operative field (0.5μm, p = 0.01; 0.7μm, p = 0.008; 1μm, p = 0.007; 2.5 μm, p = 0.006; 5um, p = 0.01and 10μm, p = 0.01). The particle counts between the OR doors were decreased by 2–42% in the Intervention group compared to the Control group, and the difference was significant for particles sized 0.5μm, 0.7μm, and 1μm (0.5μm, p=0.03, 0.7μm, p=0.02 and 1μm, p=0.03). A significant decrease (56–78%) was found in the average particle count in the Intervention group between the first 15 minutes and the last 15 minutes of the case compared to the Control group (0.5μm, p=0.01; 0.7μm, p=0.004; 1μm, p=0.002; 2.5μm, p=0.001; 5μm, p=0.004 and 10μm, p=0.01). The decrease in door openings and particle counts attained in the Intervention group were sustained over the entire study period. This study demonstrated that the use of traffic cameras was an effective and sustainable method to limit OR traffic during arthroplasty surgery. The use of door posters and traffic cameras significantly limited OR traffic and reduced door openings, which resulted in a sustained reduction in particle counts near the OR doors, and more importantly within the operative field.

Orthopaedic ProceedingsVol. 107-B(SUPP_10)
McGill University Health Centre (CA)
Openalex Percentile: Top 60%
Healthcare and Environmental Waste Management
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.