Failure of femoral neck system in Pauwels type III fracture: a fractography and finite element study

• Four FNS implants failed in Pauwels type III fractures within 2–3 months post fixation. • Fractographic and subject-specific FE analyses examined failure mechanisms. • Fractography showed ductile–brittle features in two cases and fatigue in other two. • High stress occurred in cortical screws near the side plate, coinciding failure sites. The femoral neck system (FNS) is an internal fixation device for treating femoral neck fractures. We observed four instances of single-hole FNS failure in Pauwels type III fractures within two to three post-operative months. In each case, the cortical screw failed near the side plate. The fractured surfaces were analysed using scanning electron microscopy, and subject-specific finite element (FE) models were developed from preoperative CT scans to investigate stresses and strains in implants and bones to assess the failure risks and their locations. While cases 1 and 2 exhibited a combination of ductile and brittle failure, with leafy-like structures and cleavage facets, respectively, cases 3 and 4 showed fatigue failure, evident from parallel striations. For all cases, β-phase particles were dispersed within an α-phase matrix. The FE simulations confirmed elevated stresses in the cortical screws near the single-hole side plate, correlating with the clinically observed failure. These results indicate that single-hole FNS might not offer adequate stability for Pauwels type III fractures. The fractographic and FE analysis performed in the current study provides valuable insights into the failure mechanisms of the FNS and highlighting the need for improved implant design for enhanced stability and better clinical outcomes.

Authors

Institutions

Publication Details

Journal
Materials & Design
Published
2025-10-22
DOI
https://doi.org/10.1016/j.matdes.2025.114964
Citations
1
Primary Topic
Hip and Femur Fractures
Type
article
Field-Weighted Citation Impact
2.01
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Failure of femoral neck system in Pauwels type III fracture: a fractography and finite element study

Arpan Upadhyay, Samarth Mittal, Kaushik Mukherjee, Prithviraj Mukhopadhyay et al.
1 citations
Materials & Design
Hip and Femur Fractures
2.01
article

Failure of femoral neck system in Pauwels type III fracture: a fractography and finite element study

Arpan Upadhyay, Samarth Mittal, Kaushik Mukherjee, Prithviraj Mukhopadhyay, Anoop Chawla, Vivek Trikha, L.P. Rao
article en
1 citations

Abstract

• Four FNS implants failed in Pauwels type III fractures within 2–3 months post fixation. • Fractographic and subject-specific FE analyses examined failure mechanisms. • Fractography showed ductile–brittle features in two cases and fatigue in other two. • High stress occurred in cortical screws near the side plate, coinciding failure sites. The femoral neck system (FNS) is an internal fixation device for treating femoral neck fractures. We observed four instances of single-hole FNS failure in Pauwels type III fractures within two to three post-operative months. In each case, the cortical screw failed near the side plate. The fractured surfaces were analysed using scanning electron microscopy, and subject-specific finite element (FE) models were developed from preoperative CT scans to investigate stresses and strains in implants and bones to assess the failure risks and their locations. While cases 1 and 2 exhibited a combination of ductile and brittle failure, with leafy-like structures and cleavage facets, respectively, cases 3 and 4 showed fatigue failure, evident from parallel striations. For all cases, β-phase particles were dispersed within an α-phase matrix. The FE simulations confirmed elevated stresses in the cortical screws near the single-hole side plate, correlating with the clinically observed failure. These results indicate that single-hole FNS might not offer adequate stability for Pauwels type III fractures. The fractographic and FE analysis performed in the current study provides valuable insights into the failure mechanisms of the FNS and highlighting the need for improved implant design for enhanced stability and better clinical outcomes.

Materials & DesignVol. 260
All India Institute of Medical Sciences (IN), Indian Institute of Technology Delhi (IN)
Openalex Percentile: Top 11%
Hip and Femur Fractures
2.01
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.