Exploring sevoflurane consumption and CO2 emissions of individual patients undergoing noncardiac surgery using a target-controlled sevoflurane administration system

Reduce sevoflurane consumption during anaesthesia remains an economic and environmental challenge. This case study analyzed retrospectively a large cohort of procedures using end-tidal (ET) Control to optimize sevoflurane consumption and assess the impact of ventilator settings on it. In single center, consecutive adult procedures for noncardiac surgery under general anaesthesia were analyzed in twelve operating rooms. The anaesthesia system (Aisys CS<sup>2</sup>, GE, USA) was connected to software (Carestation Insight, GE, USA) that automatically recorded sevoflurane consumption for each case. Left to the discretion of the anaesthesiologist, fresh gas flow (FGF) was set at 0.5, 0.8 or 1 L.min<sup>-1</sup> with an initial end-tidal sevoflurane target of 1.2-2% to reach the goal of approximately 1.0 MAC. The primary endpoint was the sevoflurane consumption (mL.min<sup>-1</sup>). Secondary endpoints were sevoflurane consumption and carbon footprint for the cohort (sevoflurane GWP<sub>20</sub> = 702) and by anaesthesia duration, type of airway management (endotracheal intubation EI, laryngeal mask LM), FGF and initial sevoflurane settings. From May to September 2024, 3064 procedures were recorded via the app. with a median (IQR) duration of surgery of 79 (47-124) minutes. Sevoflurane consumption was. 19 528 mL. Median (IQR) sevoflurane consumption was 0.16 (0.12-0.20) mL.min<sup>-1</sup> with a carbon footprint (sevoflurane GWP<sub>20</sub> including manufacturing = 1,468 kg/mL) of 0,23(0.17-0.29) kgCO<sub>2eq</sub>.min<sup>-1</sup>.Subgroup analysis demonstrated that FGF at 0.8 or 1.0 L.min<sup>-1</sup> significantly increased sevoflurane consumption and CO<sub>2</sub> emission when compared with 0.5 L.min<sup>-1</sup> (P < 0.01). Initial ET target sevoflurane levels of 1.2% vs greater than 1.2% did not change total sevoflurane consumption. Regarding airway management (EI vs LM), LM use was associated with higher consumption, specifically when duration > 120 min (P < 0.01). In this large cohort of cases, this study demonstrated that controlling ET with an FGF target of 0.5 L.min<sup>-1</sup> remained the best way to reduce sevoflurane consumption. Other settings did not significantly reduce gas consumption.

Authors

Institutions

Publication Details

Journal
Journal of Clinical Monitoring and Computing
Published
2025-10-21
DOI
https://doi.org/10.1007/s10877-025-01370-3
Primary Topic
Climate Change and Health Impacts
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Exploring sevoflurane consumption and CO2 emissions of individual patients undergoing noncardiac surgery using a target-controlled sevoflurane administration system

Patrice Forget, Yann Gricourt, Virginie Chasseigne, Philippe Cuvillon et al.
Journal of Clinical Monitoring and Computing
Climate Change and Health Impacts
article

Exploring sevoflurane consumption and CO2 emissions of individual patients undergoing noncardiac surgery using a target-controlled sevoflurane administration system

Patrice Forget, Yann Gricourt, Virginie Chasseigne, Philippe Cuvillon, Mikael Perin, Myriam Mezzarobba, Pierre-Baptiste Vialatte, Brenton Alexander, Jean Yves Lefrant, Guillaume Bibollet
article en

Abstract

Reduce sevoflurane consumption during anaesthesia remains an economic and environmental challenge. This case study analyzed retrospectively a large cohort of procedures using end-tidal (ET) Control to optimize sevoflurane consumption and assess the impact of ventilator settings on it. In single center, consecutive adult procedures for noncardiac surgery under general anaesthesia were analyzed in twelve operating rooms. The anaesthesia system (Aisys CS<sup>2</sup>, GE, USA) was connected to software (Carestation Insight, GE, USA) that automatically recorded sevoflurane consumption for each case. Left to the discretion of the anaesthesiologist, fresh gas flow (FGF) was set at 0.5, 0.8 or 1 L.min<sup>-1</sup> with an initial end-tidal sevoflurane target of 1.2-2% to reach the goal of approximately 1.0 MAC. The primary endpoint was the sevoflurane consumption (mL.min<sup>-1</sup>). Secondary endpoints were sevoflurane consumption and carbon footprint for the cohort (sevoflurane GWP<sub>20</sub> = 702) and by anaesthesia duration, type of airway management (endotracheal intubation EI, laryngeal mask LM), FGF and initial sevoflurane settings. From May to September 2024, 3064 procedures were recorded via the app. with a median (IQR) duration of surgery of 79 (47-124) minutes. Sevoflurane consumption was. 19 528 mL. Median (IQR) sevoflurane consumption was 0.16 (0.12-0.20) mL.min<sup>-1</sup> with a carbon footprint (sevoflurane GWP<sub>20</sub> including manufacturing = 1,468 kg/mL) of 0,23(0.17-0.29) kgCO<sub>2eq</sub>.min<sup>-1</sup>.Subgroup analysis demonstrated that FGF at 0.8 or 1.0 L.min<sup>-1</sup> significantly increased sevoflurane consumption and CO<sub>2</sub> emission when compared with 0.5 L.min<sup>-1</sup> (P < 0.01). Initial ET target sevoflurane levels of 1.2% vs greater than 1.2% did not change total sevoflurane consumption. Regarding airway management (EI vs LM), LM use was associated with higher consumption, specifically when duration > 120 min (P < 0.01). In this large cohort of cases, this study demonstrated that controlling ET with an FGF target of 0.5 L.min<sup>-1</sup> remained the best way to reduce sevoflurane consumption. Other settings did not significantly reduce gas consumption.

Journal of Clinical Monitoring and Computing
Inserm (FR), Université de Montpellier (FR), University of California San Diego (US), NHS Grampian (GB), Université de Nîmes (FR)
Openalex Percentile: Top 61%
Climate Change and Health Impacts
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.