Effect of Abutment Screw Design on Torque Loss Under Cyclic Fatigue Loading: A Comparison of TSIII and KSIII Implant Systems
The KSIII system demonstrated superior biomechanical stability, with significantly lower torque loss compared with TSIII, independent of torque level. Clinically, these findings suggest that the KSIII system may reduce the incidence of screw loosening and associated complications. A tightening torque of approximately 35 Ncm appeared to provide the most stable preload. Long-term in vivo studies are warranted to confirm these results under clinical conditions.