Detector response and dose calculation lateral to material interfaces for 6 MV photon beam
Dose calculation near inhomogeneities depends on the used algorithm, dose measurements in the same region differ depending on the detector type used. We recommend verification of dose calculation with second type of algorithm and measurements with at least two detector types.
Authors
- Otto A. Sauer (ORCID: https://orcid.org/0000-0002-4418-9779)
- Sonja Wegener (ORCID: https://orcid.org/0000-0002-3024-7069)
- Jonas Ringholz
Institutions
- Universitätsklinikum Würzburg (DE)
Publication Details
- Journal
- Journal of Applied Clinical Medical Physics
- Published
- 2025-10-21
- DOI
- https://doi.org/10.1002/acm2.70302
- Primary Topic
- Advanced Radiotherapy Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- Deutsche Forschungsgemeinschaft
- Julius-Maximilians-Universität Würzburg