Impact of the Sensor Electrical Noise on Ionoacoustic Dose Reconstruction using 200 MeV Clinical Proton Beam

Ionoacoustic dosimetry has emerged as a promising technique for real-time dose verification in proton therapy, where conventional methods like Electronic Portal Imaging Devices (EPIDs) are ineffective due to the absence of an exit dose beyond the Bragg peak. This study evaluates the impact of realistic sensor noise on the ionoacoustic reconstruction of dose distributions generated by a 200 MeV proton beam, simulated using the k -Wave toolbox. Different signal-to-noise ratios (SNRs) were modeled by varying the delivered dose per pulse across a range of clinically relevant values ($1-100 \mathrm{mGy}$). The quality of the reconstructed dose distributions was assessed using gamma index analysis at $1 \% / 1$ mm and $3 \% / 3 \mathrm{~mm}$ criteria. Results demonstrate that, even under significant noise conditions, accurate dose reconstructions are achievable without the need for temporal averaging, provided sufficient per-pulse dose. These findings support the feasibility of ionoacoustic dosimetry for online, high-precision proton therapy monitoring and suggest pathways for future experimental validation and clinical translation.

Authors

Institutions

Publication Details

Published
2025-09-21
DOI
https://doi.org/10.1109/prime66228.2025.11203461
Primary Topic
Radiation Therapy and Dosimetry
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Impact of the Sensor Electrical Noise on Ionoacoustic Dose Reconstruction using 200 MeV Clinical Proton Beam

Marcello De Matteis, Mara Severgnini, A. Basçhirotto, Maurizio Marrale et al.
Radiation Therapy and Dosimetry
article

Impact of the Sensor Electrical Noise on Ionoacoustic Dose Reconstruction using 200 MeV Clinical Proton Beam

Marcello De Matteis, Mara Severgnini, A. Basçhirotto, Maurizio Marrale, Elia A. Vallicelli, Alessandro Ferrara, Saadat Ali
article en

Abstract

Ionoacoustic dosimetry has emerged as a promising technique for real-time dose verification in proton therapy, where conventional methods like Electronic Portal Imaging Devices (EPIDs) are ineffective due to the absence of an exit dose beyond the Bragg peak. This study evaluates the impact of realistic sensor noise on the ionoacoustic reconstruction of dose distributions generated by a 200 MeV proton beam, simulated using the k -Wave toolbox. Different signal-to-noise ratios (SNRs) were modeled by varying the delivered dose per pulse across a range of clinically relevant values ($1-100 \mathrm{mGy}$). The quality of the reconstructed dose distributions was assessed using gamma index analysis at $1 \% / 1$ mm and $3 \% / 3 \mathrm{~mm}$ criteria. Results demonstrate that, even under significant noise conditions, accurate dose reconstructions are achievable without the need for temporal averaging, provided sufficient per-pulse dose. These findings support the feasibility of ionoacoustic dosimetry for online, high-precision proton therapy monitoring and suggest pathways for future experimental validation and clinical translation.

University of Padua (IT), Istituto Universitario di Studi Superiori di Pavia (IT), Istituto Nazionale di Fisica Nucleare, Sezione di Milano Bicocca (IT), Azienda Sanitaria Universitaria Integrata di Trieste (IT), University of Milano-Bicocca (IT), University of Palermo (IT)
Openalex Percentile: Top 57%
Radiation Therapy and Dosimetry
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.