High Frame Rate Arterial Monitoring via Wi-Fi 6 on a 32-channel Wearable Ultrasound Probe

Cardiovascular monitoring (CM) is an exemplary high-impact scenario for wearable ultrasound (WUS), and multichannel WUS is needed to enable accurate vessel scanning and dynamic optimal channel selection for vessel diameter extraction. However, current WUS probes remain bottlenecked by their wireless link, which significantly impacts power consumption and bounds frame rates (FR) typically to less than 50 Hz, limiting their effectiveness in CM, especially for tracking fast dynamics in high beats per minute (bpm) scenarios. To overcome these limitations, we design a novel wireless solution using Wi-Fi 6 to improve energy efficiency and enable FR as high as 175 Hz on a 32-channel WUS probe, demonstrating its use for CM at high pulse rates (130 bpm). The Wi-Fi solution achieves a 35.7 Mbps wireless throughput with 486 mW total link power consumption and a transmission efficiency of 73.5 Mbps/W, 46% higher than Wi-Fi 4-based systems.

Authors

Publication Details

Published
2025-09-15
DOI
https://doi.org/10.1109/ius62464.2025.11201780
Primary Topic
Non-Invasive Vital Sign Monitoring
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

High Frame Rate Arterial Monitoring via Wi-Fi 6 on a 32-channel Wearable Ultrasound Probe

Luca Benini, Sergei Vostrikov, Andrea Cossettini, Cédric Hirschi
Non-Invasive Vital Sign Monitoring
article

High Frame Rate Arterial Monitoring via Wi-Fi 6 on a 32-channel Wearable Ultrasound Probe

Luca Benini, Sergei Vostrikov, Andrea Cossettini, Cédric Hirschi
article en

Abstract

Cardiovascular monitoring (CM) is an exemplary high-impact scenario for wearable ultrasound (WUS), and multichannel WUS is needed to enable accurate vessel scanning and dynamic optimal channel selection for vessel diameter extraction. However, current WUS probes remain bottlenecked by their wireless link, which significantly impacts power consumption and bounds frame rates (FR) typically to less than 50 Hz, limiting their effectiveness in CM, especially for tracking fast dynamics in high beats per minute (bpm) scenarios. To overcome these limitations, we design a novel wireless solution using Wi-Fi 6 to improve energy efficiency and enable FR as high as 175 Hz on a 32-channel WUS probe, demonstrating its use for CM at high pulse rates (130 bpm). The Wi-Fi solution achieves a 35.7 Mbps wireless throughput with 486 mW total link power consumption and a transmission efficiency of 73.5 Mbps/W, 46% higher than Wi-Fi 4-based systems.

Openalex Percentile: Top 71%
Non-Invasive Vital Sign Monitoring
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.