Analysis of Heme Using a Combination of Total Reflection X‐Ray Fluorescence Analysis and Solid‐Phase Extraction

ABSTRACT In this study, a method to analyze iron originating from heme using a combination of total reflection X‐ray fluorescence (TXRF) analysis and solid‐phase extraction is presented. In this method, heme in a sample solution was collected on a reverse‐phase solid‐phase extraction column filled with octadecyl bonded silica, and then the collected heme was eluted with an eluent. Finally, the dry residue of a 10 μL droplet of the eluate diluted with ultrapure water was measured by a portable TXRF spectrometer. The Fe Kα peak originating from heme was observed by using this combination, and the detection limit for heme was estimated to be about 1 μmol L −1 .

Authors

Institutions

Publication Details

Journal
X-Ray Spectrometry
Published
2025-10-19
DOI
https://doi.org/10.1002/xrs.70049
Primary Topic
Porphyrin Metabolism and Disorders
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Analysis of Heme Using a Combination of Total Reflection X‐Ray Fluorescence Analysis and Solid‐Phase Extraction

Hajime Hoshi, Shinsuke Kunimura
X-Ray Spectrometry
Porphyrin Metabolism and Disorders
article

Analysis of Heme Using a Combination of Total Reflection X‐Ray Fluorescence Analysis and Solid‐Phase Extraction

Hajime Hoshi, Shinsuke Kunimura
article en

Abstract

ABSTRACT In this study, a method to analyze iron originating from heme using a combination of total reflection X‐ray fluorescence (TXRF) analysis and solid‐phase extraction is presented. In this method, heme in a sample solution was collected on a reverse‐phase solid‐phase extraction column filled with octadecyl bonded silica, and then the collected heme was eluted with an eluent. Finally, the dry residue of a 10 μL droplet of the eluate diluted with ultrapure water was measured by a portable TXRF spectrometer. The Fe Kα peak originating from heme was observed by using this combination, and the detection limit for heme was estimated to be about 1 μmol L −1 .

X-Ray SpectrometryVol. 55(2)
Tokyo University of Science (JP), Advanced Institute of Industrial Technology (JP)
Openalex Percentile: Top 74%
Porphyrin Metabolism and Disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.