Effect of Walking Speed on the Reliability of a Smartphone-Based Markerless Gait Analysis System

Quantitative gait analysis is essential for understanding motor function and guiding clinical decisions. While marker-based motion capture (MoCap) systems are accurate, they are costly and require specialized facilities. OpenCap, a markerless alternative, offers a more accessible approach; however, its reliability across different walking speeds remains uncertain. This study assessed the agreement between OpenCap and MoCap in measuring spatiotemporal parameters, joint kinematics, and center of mass (CoM) displacement during level walking at three speeds: slow, self-selected, and fast. Fifteen healthy adults performed multiple trials simultaneously, recorded by both systems. Agreement was analyzed using intraclass correlation coefficients (ICC), minimal detectable change (MDC), Bland-Altman analyses, root mean square error (RMSE), Statistical Parametric Mapping (SPM), and repeated-measures ANOVA. Results indicated excellent agreement for spatiotemporal variables (ICC ≥ 0.95) and high consistency for joint waveforms (RMSE < 2°) and CoM displacement (RMSE < 6 mm) across all speeds. However, the joint range of motion (ROM) showed lower reliability, especially at the hip and ankle, at higher speeds. ANOVA revealed no significant System × Speed interactions for most variables, though a significant effect of speed was noted, with OpenCap underestimating walking speed more at fast speeds. Overall, OpenCap is a valuable tool for gait assessment, very accurate for spatiotemporal data and CoM displacement. Still, caution should be taken when interpreting joint kinematics and speed at different walking speeds.

Authors

Institutions

Publication Details

Journal
Sensors
Published
2025-10-20
DOI
https://doi.org/10.3390/s25206474
Citations
2
Primary Topic
Muscle activation and electromyography studies
Type
article
Field-Weighted Citation Impact
1.35
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Effect of Walking Speed on the Reliability of a Smartphone-Based Markerless Gait Analysis System

Paolo Capodaglio, Verônica Cimolin, Marcus Peikriszwili Tartaruga, Leonardo Alexandre Peyré‐Tartaruga et al.
2 citations
Sensors
Muscle activation and electromyography studies
1.35
article

Effect of Walking Speed on the Reliability of a Smartphone-Based Markerless Gait Analysis System

Paolo Capodaglio, Verônica Cimolin, Marcus Peikriszwili Tartaruga, Leonardo Alexandre Peyré‐Tartaruga, Edilson Fernando de Borba, Serena Cerfoglio, Jorge L. Storniolo, Paolo Cavallari
article en
2 citations

Abstract

Quantitative gait analysis is essential for understanding motor function and guiding clinical decisions. While marker-based motion capture (MoCap) systems are accurate, they are costly and require specialized facilities. OpenCap, a markerless alternative, offers a more accessible approach; however, its reliability across different walking speeds remains uncertain. This study assessed the agreement between OpenCap and MoCap in measuring spatiotemporal parameters, joint kinematics, and center of mass (CoM) displacement during level walking at three speeds: slow, self-selected, and fast. Fifteen healthy adults performed multiple trials simultaneously, recorded by both systems. Agreement was analyzed using intraclass correlation coefficients (ICC), minimal detectable change (MDC), Bland-Altman analyses, root mean square error (RMSE), Statistical Parametric Mapping (SPM), and repeated-measures ANOVA. Results indicated excellent agreement for spatiotemporal variables (ICC ≥ 0.95) and high consistency for joint waveforms (RMSE < 2°) and CoM displacement (RMSE < 6 mm) across all speeds. However, the joint range of motion (ROM) showed lower reliability, especially at the hip and ankle, at higher speeds. ANOVA revealed no significant System × Speed interactions for most variables, though a significant effect of speed was noted, with OpenCap underestimating walking speed more at fast speeds. Overall, OpenCap is a valuable tool for gait assessment, very accurate for spatiotemporal data and CoM displacement. Still, caution should be taken when interpreting joint kinematics and speed at different walking speeds.

SensorsVol. 25(20)
University of Milan (IT), University of Pavia (IT), Universidade Estadual do Centro-Oeste (BR), IRCCS Istituto Auxologico Italiano (IT), Midwestern State University (US), Universidade Federal do Paraná (BR), Politecnico di Milano (IT)
Openalex Percentile: Top 17%
Muscle activation and electromyography studies
1.35
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.