Point-of-Care Solid-Phase PCR in a Vertical Microfluidic Chip Integrated with All-Dielectric Nanostructured Metasurface for Highly Sensitive, Multiplexed Pathogen Detection
Multiplexed solid-phase polymerase chain reaction (SP-PCR) has emerged as an indispensable modality for concurrent amplification of multiple genetic loci within a singular reaction vessel, facilitating efficient molecular diagnostics. Nevertheless, SP-PCR has seldom been integrated into point-of-care diagnostic devices due to several technical challenges, such as bubble formation during PCR, long reaction time, and low fluorescence signals generated from the PCR products on a solid surface. To circumvent these constraints, we engineered a microfluidic chip comprising SP-PCR and nanophotonic enhancement to enable highly sensitive, high-throughput, and cost-efficient molecular diagnostics. The chip's vertical orientation integrates preloaded reagent chambers for sequential lysis, washing, elution, and amplification, driven by a synchronized stepper motor and air vacuum, achieving robust nucleic acid purification and reverse transcription-PCR, and enabling bubble-free, gravity-assisted fluid dynamics during the PCR thermocycling. Thermal cycling is expedited through a dual-heater configuration alternating at subsecond intervals, obviating active cooling and shortening the reaction time. All-dielectric nanostructured metasurface was incorporated beneath the PCR chamber, allowing for the facile immobilization of DNA arrays to conduct SP-PCR. Taking advantage of guided-mode resonance supported by the metasurface and the SP-PCR approaches permits multiplexed detection and achieves a detection limit of 10 copies/reaction, highlighting the platform's potential for point-of-care diagnostics, personalized medicine, and high-throughput pathogen surveillance. Facile fabrication and automation emphasize scalability for mass production and deployment and collectively represent an advancement in point-of-care diagnostics.
Authors
- Jan Gorodkin (ORCID: https://orcid.org/0000-0001-5823-4000)
- Leonid Yu. Beliaev (ORCID: https://orcid.org/0000-0002-5695-4896)
- Christian Anthon (ORCID: https://orcid.org/0000-0002-4336-1688)
- Sanshui Xiao (ORCID: https://orcid.org/0000-0001-6529-5047)
- Islam Seder (ORCID: https://orcid.org/0000-0001-7204-9741)
- Yi Sun (ORCID: https://orcid.org/0000-0002-0210-4730)
- Tao Zheng (ORCID: https://orcid.org/0000-0002-0626-4344)
- Rodrigo Coronel Téllez
- Dhouha Grissa
Institutions
- University of Copenhagen (DK)
- Ørsted (Denmark) (DK)
- NeoPhotonics (United States) (US)
- Technical University of Denmark (DK)
Publication Details
- Journal
- ACS Sensors
- Published
- 2025-10-19
- DOI
- https://doi.org/10.1021/acssensors.5c02435
- Citations
- 2
- Primary Topic
- Biosensors and Analytical Detection
- Type
- article
- Field-Weighted Citation Impact
- 1.35