A study of multi-parameter impact on beam position accuracy for scanning proton therapy

Pencil beam scanning (PBS) has become the dominant beam delivery method in proton therapy facilities, owing to its high dose conformality and high beam utilization. While prescribed dose distribution necessitates accurate control of both irradiation position and dose, critical factors including magnetic field fluctuation, mechanical deformation, and beam orbit distortion may impact irradiation position accuracy. This study quantitatively evaluates the impacts of these three factors on irradiation position accuracy for the SC200 proton therapy system. Firstly, three-dimensional magnetic field data of nozzle were acquired through COMSOL simulation and subsequently validated against experimental measurements. Secondly, the simulated field data were utilized to Monte Carlo assessment of individual parameter effects: (1) power supply ripple and (2) beam orbit distortion on irradiation position accuracy, along with (3) analysis of gantry deformation-induced isocenter shifts. Thirdly, the combined effects of these three critical parameters on irradiation accuracy were systematically analyzed. Results reveal that irradiation position deviation patterns under multi-parameter interactions exhibit complex behavior, with beam orbit fluctuations showing the most significant impact in three-dimensional magnetic field, particularly through quadrant-dependent deviation inconsistencies. The study further proposes orienting beam orbit distortion toward the fourth quadrant, demonstrating that this approach can largely maintain irradiation position deviations within ±0.5 mm in both X and Y directions under combined parameter effects. These findings establish a foundation for future high-precision dose delivery system design.

Authors

Institutions

Publication Details

Journal
Nuclear Engineering and Technology
Published
2025-10-17
DOI
https://doi.org/10.1016/j.net.2025.103989
Primary Topic
Radiation Therapy and Dosimetry
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A study of multi-parameter impact on beam position accuracy for scanning proton therapy

Junsong Shen, Ming Wang, Xianhu Zeng, Guo-Yu Xie et al.
Nuclear Engineering and Technology
Radiation Therapy and Dosimetry
article

A study of multi-parameter impact on beam position accuracy for scanning proton therapy

Junsong Shen, Ming Wang, Xianhu Zeng, Guo-Yu Xie, Sheng Liao
article en

Abstract

Pencil beam scanning (PBS) has become the dominant beam delivery method in proton therapy facilities, owing to its high dose conformality and high beam utilization. While prescribed dose distribution necessitates accurate control of both irradiation position and dose, critical factors including magnetic field fluctuation, mechanical deformation, and beam orbit distortion may impact irradiation position accuracy. This study quantitatively evaluates the impacts of these three factors on irradiation position accuracy for the SC200 proton therapy system. Firstly, three-dimensional magnetic field data of nozzle were acquired through COMSOL simulation and subsequently validated against experimental measurements. Secondly, the simulated field data were utilized to Monte Carlo assessment of individual parameter effects: (1) power supply ripple and (2) beam orbit distortion on irradiation position accuracy, along with (3) analysis of gantry deformation-induced isocenter shifts. Thirdly, the combined effects of these three critical parameters on irradiation accuracy were systematically analyzed. Results reveal that irradiation position deviation patterns under multi-parameter interactions exhibit complex behavior, with beam orbit fluctuations showing the most significant impact in three-dimensional magnetic field, particularly through quadrant-dependent deviation inconsistencies. The study further proposes orienting beam orbit distortion toward the fourth quadrant, demonstrating that this approach can largely maintain irradiation position deviations within ±0.5 mm in both X and Y directions under combined parameter effects. These findings establish a foundation for future high-precision dose delivery system design.

Nuclear Engineering and TechnologyVol. 58(2)
Chinese Academy of Sciences (CN), Chengdu University of Technology (CN), Hefei CAS Ion Medical and Technical Devices (China) (CN), Institute of Plasma Physics (CN)
Openalex Percentile: Top 58%
Radiation Therapy and Dosimetry
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.