Reducing False Alarm Rates and Workload in ICUs by Improving Arrhythmia Detection Algorithms of Patient Monitoring Systems

BACKGROUND Patient monitoring systems are widely used in intensive care units (ICUs) to monitor patient's conditions. A high false alarm rate can lead to alarm fatigue among nurses, increasing workload and stress. This study aimed to improve the accuracy of arrhythmia detection by enhancing the noise detection algorithm in patient monitoring systems and to determine whether false alarm rate and workload decreased through clinical trials. MATERIAL AND METHODS Trials were conducted on adult patients in the ICU at Yongin Severance Hospital who required continuous electrocardiogram (ECG) monitoring for at least 2 days. After the first trial, the noise detection algorithm of the M50 (investigational device) was improved, and a second trial was conducted to evaluate its performance. Both trials followed the same study design. During the study period, M50 and MX700 (comparator device) were applied simultaneously for 3 days. Arrhythmia alarms were reviewed by an independent evaluator who assessed false alarms by comparing them with the ECG signals. False alarm rates were compared between trials using the chi-square (χ²) test. RESULTS The clinical trial was conducted through 2 separate trials, with 17 and 11 participants, respectively. A comparative analysis of false alarm rates of the investigational device demonstrated a reduction from 71.75% to 27.61%. Statistical analysis using the chi-square test indicated a P value of 0.000 (<0.001), confirming a statistically significant difference. CONCLUSIONS The results of 2 trials demonstrated reductions in false alarm rate and NASA-TLX score. These findings suggest that enhancing the noise detection algorithm in the patient monitoring system improved arrhythmia detection accuracy and helped reduce nurses' workload.

Authors

Institutions

Publication Details

Journal
Medical Science Monitor
Published
2025-10-17
DOI
https://doi.org/10.12659/msm.949932
Primary Topic
Healthcare Technology and Patient Monitoring
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Reducing False Alarm Rates and Workload in ICUs by Improving Arrhythmia Detection Algorithms of Patient Monitoring Systems

Jae‐Sun Uhm, Yong Ju Lee, Ji Soo Choi, Jiyoon Oh et al.
Medical Science Monitor
Healthcare Technology and Patient Monitoring
article

Reducing False Alarm Rates and Workload in ICUs by Improving Arrhythmia Detection Algorithms of Patient Monitoring Systems

Jae‐Sun Uhm, Yong Ju Lee, Ji Soo Choi, Jiyoon Oh, Won Seuk Jang, Chul‐Min Ahn, Youngjae Kim, Dae Keun Jeon
article en

Abstract

BACKGROUND Patient monitoring systems are widely used in intensive care units (ICUs) to monitor patient's conditions. A high false alarm rate can lead to alarm fatigue among nurses, increasing workload and stress. This study aimed to improve the accuracy of arrhythmia detection by enhancing the noise detection algorithm in patient monitoring systems and to determine whether false alarm rate and workload decreased through clinical trials. MATERIAL AND METHODS Trials were conducted on adult patients in the ICU at Yongin Severance Hospital who required continuous electrocardiogram (ECG) monitoring for at least 2 days. After the first trial, the noise detection algorithm of the M50 (investigational device) was improved, and a second trial was conducted to evaluate its performance. Both trials followed the same study design. During the study period, M50 and MX700 (comparator device) were applied simultaneously for 3 days. Arrhythmia alarms were reviewed by an independent evaluator who assessed false alarms by comparing them with the ECG signals. False alarm rates were compared between trials using the chi-square (χ²) test. RESULTS The clinical trial was conducted through 2 separate trials, with 17 and 11 participants, respectively. A comparative analysis of false alarm rates of the investigational device demonstrated a reduction from 71.75% to 27.61%. Statistical analysis using the chi-square test indicated a P value of 0.000 (<0.001), confirming a statistically significant difference. CONCLUSIONS The results of 2 trials demonstrated reductions in false alarm rate and NASA-TLX score. These findings suggest that enhancing the noise detection algorithm in the patient monitoring system improved arrhythmia detection accuracy and helped reduce nurses' workload.

Medical Science MonitorVol. 31
Yonsei University (KR), Severance Hospital (KR), Gangnam Severance Hospital (KR), Yonsei University Health System (KR), Seoul Media Institute of Technology (KR)
Openalex Percentile: Top 54%
Healthcare Technology and Patient Monitoring
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.