Evaluation of Emerging Technologies to Aid in the Detection and Diagnosis of Acute Extremity Compartment Syndrome

<b>Background/Objectives</b>: The diagnosis of acute compartment syndrome (ACS) of the extremities is typically based on subjective clinical signs and symptoms, highlighting the need for user-friendly diagnostic tools to improve accuracy and reliability. This study evaluates the performance of two commercial devices, the MY01<sup>®</sup> continuous pressure monitoring system and the Moxy Monitor near-infrared spectroscopy-based system, against a reference standard of continuous intracompartmental pressure (ICP) monitoring in a preclinical ACS model. <b>Methods</b>: ACS was induced in the anterior compartment of the distal hind limb in eight Yorkshire pigs using a balloon displacement model. ICP was incrementally elevated and maintained for four hours at >30 mmHg above mean arterial pressure. This was followed by balloon deflation and reperfusion. Final assessments were performed at 24 h post-injury. ICP measurements from the MY01<sup>®</sup> and muscle oxygen saturation (SmO<sub>2</sub>) data from the Moxy Monitor were compared to reference ICP measurements. Histologic analysis of muscle tissue was performed to assess the severity of necrosis. <b>Results</b>: The MY01<sup>®</sup> provided accurate ICP measurements, with a mean bias of 2.21 ± 18.77 mmHg during pre-ischemia, 4.86 ± 10.43 mmHg during reperfusion, and 4.69 ± 3.28 mmHg 24 h post-injury, compared to reference probes. Correlation at 24 h post-injury was (r = 0.86, R<sup>2</sup> = 0.73, <i>p</i> < 0.0001). In contrast, the Moxy Monitor failed to detect significant differences in SmO<sub>2</sub> between injured and control limbs at 24 h post-injury, despite pronounced ICP differences. Our volumetric displacement ACS model demonstrated its efficacy as a testing platform by allowing for controlled, incremental elevation in ICP and sustaining elevated ICP levels after 24 h. Histologic evaluation confirmed extensive muscle damage, including edema and necrosis. <b>Conclusions</b>: The MY01<sup>®</sup> provides accurate, continuous ICP monitoring, supporting its clinical utility in ACS diagnosis. However, the use of near-infrared spectroscopy-based systems such as the Moxy Monitor for ACS diagnosis and management should continue to be critically scrutinized.

Authors

Institutions

Publication Details

Journal
Diagnostics
Published
2025-10-16
DOI
https://doi.org/10.3390/diagnostics15202607
Primary Topic
Muscle and Compartmental Disorders
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of Emerging Technologies to Aid in the Detection and Diagnosis of Acute Extremity Compartment Syndrome

Stefanie M. Shiels, R Moritz, Daniel J. Cognetti, Andréanne Powers et al.
Diagnostics
Muscle and Compartmental Disorders
article

Evaluation of Emerging Technologies to Aid in the Detection and Diagnosis of Acute Extremity Compartment Syndrome

Stefanie M. Shiels, R Moritz, Daniel J. Cognetti, Andréanne Powers, Robert V. Hainline, ThomasW. Walters, Catharina Gaeth, Kinton Armmer
article en

Abstract

<b>Background/Objectives</b>: The diagnosis of acute compartment syndrome (ACS) of the extremities is typically based on subjective clinical signs and symptoms, highlighting the need for user-friendly diagnostic tools to improve accuracy and reliability. This study evaluates the performance of two commercial devices, the MY01<sup>®</sup> continuous pressure monitoring system and the Moxy Monitor near-infrared spectroscopy-based system, against a reference standard of continuous intracompartmental pressure (ICP) monitoring in a preclinical ACS model. <b>Methods</b>: ACS was induced in the anterior compartment of the distal hind limb in eight Yorkshire pigs using a balloon displacement model. ICP was incrementally elevated and maintained for four hours at >30 mmHg above mean arterial pressure. This was followed by balloon deflation and reperfusion. Final assessments were performed at 24 h post-injury. ICP measurements from the MY01<sup>®</sup> and muscle oxygen saturation (SmO<sub>2</sub>) data from the Moxy Monitor were compared to reference ICP measurements. Histologic analysis of muscle tissue was performed to assess the severity of necrosis. <b>Results</b>: The MY01<sup>®</sup> provided accurate ICP measurements, with a mean bias of 2.21 ± 18.77 mmHg during pre-ischemia, 4.86 ± 10.43 mmHg during reperfusion, and 4.69 ± 3.28 mmHg 24 h post-injury, compared to reference probes. Correlation at 24 h post-injury was (r = 0.86, R<sup>2</sup> = 0.73, <i>p</i> < 0.0001). In contrast, the Moxy Monitor failed to detect significant differences in SmO<sub>2</sub> between injured and control limbs at 24 h post-injury, despite pronounced ICP differences. Our volumetric displacement ACS model demonstrated its efficacy as a testing platform by allowing for controlled, incremental elevation in ICP and sustaining elevated ICP levels after 24 h. Histologic evaluation confirmed extensive muscle damage, including edema and necrosis. <b>Conclusions</b>: The MY01<sup>®</sup> provides accurate, continuous ICP monitoring, supporting its clinical utility in ACS diagnosis. However, the use of near-infrared spectroscopy-based systems such as the Moxy Monitor for ACS diagnosis and management should continue to be critically scrutinized.

DiagnosticsVol. 15(20)
United States Army Institute of Surgical Research (US)
Openalex Percentile: Top 54%
Muscle and Compartmental Disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.