Single-distance, phase-only frequency-domain NIRS for vascular oxygenation and tissue metabolism: a Monte Carlo simulation study
These results validate single-separation phase-only FD-NIRS as an accurate method for multichromophore quantification (including <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"> <mml:mrow><mml:msub><mml:mi>CCO</mml:mi> <mml:mtext>redox</mml:mtext></mml:msub> </mml:mrow> </mml:math> ), enabling simpler, cost-effective systems without compromising metabolic imaging capability. The approach achieves <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"><mml:mrow><mml:mo><</mml:mo> <mml:mn>10</mml:mn> <mml:mo>%</mml:mo></mml:mrow> </mml:math> error in hemoglobin quantification while eliminating traditional multidistance requirements.