A Direct Comparison of the Performance of HemoCue Plasma/Low HB Analyzer versus Optical Spectroscopy for Quality Control Measurements of Free Hemoglobin in Blood Components

While obtained results were highly similar using both devices when assessing fHb in EC, HemoCue showed a consistent overestimation of fHb in FP samples compared to spectrophotometry. HemoCue demonstrated acceptable intraday and interday precision across concentration ranges and offers operational advantages, including faster turnaround times. Altogether in quality control analyses, HemoCue turned out to be a valuable tool for fHb measurements in EC with some limitations for FP.

Authors

Institutions

Publication Details

Journal
Transfusion Medicine and Hemotherapy
Published
2025-10-16
DOI
https://doi.org/10.1159/000548744
Primary Topic
Non-Invasive Vital Sign Monitoring
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A Direct Comparison of the Performance of HemoCue Plasma/Low HB Analyzer versus Optical Spectroscopy for Quality Control Measurements of Free Hemoglobin in Blood Components

Lisa Seekircher, Peter Willeit, Lena Tschiderer, Marco Amato et al.
Transfusion Medicine and Hemotherapy
Non-Invasive Vital Sign Monitoring
article

A Direct Comparison of the Performance of HemoCue Plasma/Low HB Analyzer versus Optical Spectroscopy for Quality Control Measurements of Free Hemoglobin in Blood Components

Lisa Seekircher, Peter Willeit, Lena Tschiderer, Marco Amato, Anita Siller, Daniela Schmidt, Harald Schennach
article en

Abstract

While obtained results were highly similar using both devices when assessing fHb in EC, HemoCue showed a consistent overestimation of fHb in FP samples compared to spectrophotometry. HemoCue demonstrated acceptable intraday and interday precision across concentration ranges and offers operational advantages, including faster turnaround times. Altogether in quality control analyses, HemoCue turned out to be a valuable tool for fHb measurements in EC with some limitations for FP.

Transfusion Medicine and Hemotherapy
Innsbruck Medical University (AT), Universität Innsbruck (AT), University of Cambridge (GB), Tirol Kliniken (AT)
Openalex Percentile: Top 72%
Non-Invasive Vital Sign Monitoring
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.