Inter-institutional variability in CT-to-mass-density conversion tables for helical tomotherapy: a national survey in Japan
This study evaluates current practices and challenges associated with computed tomography number-to-mass density (CT-MD) conversion tables in helical tomotherapy across Japan and explores directions for standardization and quality improvement amid the increasing adoption of adaptive radiotherapy (ART). A nationwide web-based survey was conducted across 34 institutions utilizing the Radixact system. Data were collected on CT acquisition protocols, calibration phantoms, density plugs, reconstruction algorithms, table registration timing and quality assurance (QA) frequency. Registered CT-MD tables were categorized by CT modality: Simulation CT (SimCT), ClearRT and CTrue. ClearRT tables were analyzed by phantom setup (full vs half), and CTrue tables by reconstruction method [filtered back projection (FPB) vs iterative reconstruction (IR)]. Inter-institutional variations in CT numbers and the number of data points were assessed. SimCT tables exhibited the widest variation in the number of data points (median = 10) and high-density CT numbers. ClearRT tables (median = 8) showed variations of up to 300 Hounsfield units (HU) in cortical bone; the half-phantom setup reduced inter-institutional variability. CTrue tables (median = 8) demonstrated high consistency, with negligible differences between IR and FPB. All plug CT numbers of CTrue remained within the tolerance defined by the American Association of Physicists in Medicine Task Group 148. However, CT numbers for air plugs varied by ~±30 HU, indicating inconsistent handling of air reference values. Additionally, 43% of institutions did not perform routine QA. Standardizing phantom geometry, air CT number handling and QA protocols-particularly using half-phantom calibration-may improve CT-MD table consistency and dose accuracy in ART.
Authors
- Hidetoshi Shimizu (ORCID: https://orcid.org/0000-0002-6633-0964)
- Hiroshi Fukuma
- Yasuo Shiota
- Takashi Hashido
- Iori Sumida
- Shuichi Ozawa
- Yuto Kitagawa
- Satoshi Kito (ORCID: https://orcid.org/0000-0002-8360-5544)
- Ryosei Nakada (ORCID: https://orcid.org/0000-0001-9238-8927)
- Keiki Sugi (ORCID: https://orcid.org/0000-0002-7608-4541)
- Shogo Tsunemine (ORCID: https://orcid.org/0009-0001-4471-0237)
- Takehiro Shiinoki (ORCID: https://orcid.org/0000-0002-0246-8489)
- Masumi Numano
- Hideyuki Harada (ORCID: https://orcid.org/0000-0001-6450-3027)
Institutions
- Mitsubishi Research Institute (Japan) (JP)
- Development Bank of Japan (JP)
- Yamaguchi University (JP)
- Tokyo Metropolitan Komagome Hospital (JP)
- Nagoya City University (JP)
- Shizuoka Cancer Center (JP)
- Iwata City Hospital (JP)
- Fujita Health University Hospital (JP)
- Hiroshima Prefectural Rehabilitation Center (JP)
- Futaba (Japan) (JP)
- The University of Osaka (JP)
Publication Details
- Journal
- Journal of Radiation Research
- Published
- 2025-09-15
- DOI
- https://doi.org/10.1093/jrr/rraf063
- Primary Topic
- Advanced Radiotherapy Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- Accuray