Significant dose reduction in diagnostic cerebral angiography using the ALARA principle and state-of-the-art medical equipment

Operator experience, the use of low frame rates during both fluoroscopy and DSA, and the application of modern medical equipment are crucial for ensuring optimal image quality while maintaining minimal exposure to ionising radiation during cerebral angiography. Our centre achieves this by using the lowest possible fluoroscopy frame rate (3.125 frames per second), a the operator set the nominal acquisition parameters at 2 fps for digital subtraction angiography (DSA), three-dimensional (3D) rotational angiography, two-dimensional /3D mapping, and roadmapping, without compromising the quality of the images obtained. A total of 213 diagnostic procedures for intracranial aneurysms, performed at a single medical centre between 2018 and 2024, were analyzed. Key exposure parameters were collected and analyzed for each procedure, including dose area product (DAP), air kerma (Ka,r), fluoroscopy time (FT), and the total number of DSA frames. The local diagnostic reference level (DRL) and typical values for cerebral angiography were determined and compared with data available in the relevant literature. The 75th percentile (P75) values, assumed to be the DRL, were 23.57 Gy · cm<sup>2</sup>for DAP (literature range: 90-209), 153 mGy for Ka,r (literature range: 711.3-1401 mGy), 10 min and 46 s for FT, and 257 for the number of DSA frames during diagnostic procedures. Typical values (50th percentile [P50]) were 13.4 Gy · cm<sup>2</sup>for DAP (literature range: 43.1-157), 92 mGy for Ka,r (literature range: 312-598 mGy), six minutes and 20 s for FT, and 182 for the number of DSA frames during diagnostic procedures. The P50 dose DRLs, measured via DAP, Ka,r, FT, and the number of DSA frames, were at least three times lower at our facility than those reported in studies on cerebral angiography published between 2013 and 2023. The use of low frame rates during both fluoroscopy and DSA has a significant impact on the radiation dose measured by DAP, Ka,r, FT, and the number of DSA frames. The proposed DRLs could be used to ensure quality and patient safety in diagnostic procedures.

Authors

Institutions

Publication Details

Journal
Journal of Radiological Protection
Published
2025-10-16
DOI
https://doi.org/10.1088/1361-6498/ae1403
Citations
1
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
1.36
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Significant dose reduction in diagnostic cerebral angiography using the ALARA principle and state-of-the-art medical equipment

M. Budzanowski, Mariusz Sowa, Kamil Węglarz, Joanna Sowa
1 citations
Journal of Radiological Protection
Radiation Dose and Imaging
1.36
article

Significant dose reduction in diagnostic cerebral angiography using the ALARA principle and state-of-the-art medical equipment

M. Budzanowski, Mariusz Sowa, Kamil Węglarz, Joanna Sowa
article en
1 citations

Abstract

Operator experience, the use of low frame rates during both fluoroscopy and DSA, and the application of modern medical equipment are crucial for ensuring optimal image quality while maintaining minimal exposure to ionising radiation during cerebral angiography. Our centre achieves this by using the lowest possible fluoroscopy frame rate (3.125 frames per second), a the operator set the nominal acquisition parameters at 2 fps for digital subtraction angiography (DSA), three-dimensional (3D) rotational angiography, two-dimensional /3D mapping, and roadmapping, without compromising the quality of the images obtained. A total of 213 diagnostic procedures for intracranial aneurysms, performed at a single medical centre between 2018 and 2024, were analyzed. Key exposure parameters were collected and analyzed for each procedure, including dose area product (DAP), air kerma (Ka,r), fluoroscopy time (FT), and the total number of DSA frames. The local diagnostic reference level (DRL) and typical values for cerebral angiography were determined and compared with data available in the relevant literature. The 75th percentile (P75) values, assumed to be the DRL, were 23.57 Gy · cm<sup>2</sup>for DAP (literature range: 90-209), 153 mGy for Ka,r (literature range: 711.3-1401 mGy), 10 min and 46 s for FT, and 257 for the number of DSA frames during diagnostic procedures. Typical values (50th percentile [P50]) were 13.4 Gy · cm<sup>2</sup>for DAP (literature range: 43.1-157), 92 mGy for Ka,r (literature range: 312-598 mGy), six minutes and 20 s for FT, and 182 for the number of DSA frames during diagnostic procedures. The P50 dose DRLs, measured via DAP, Ka,r, FT, and the number of DSA frames, were at least three times lower at our facility than those reported in studies on cerebral angiography published between 2013 and 2023. The use of low frame rates during both fluoroscopy and DSA has a significant impact on the radiation dose measured by DAP, Ka,r, FT, and the number of DSA frames. The proposed DRLs could be used to ensure quality and patient safety in diagnostic procedures.

Journal of Radiological ProtectionVol. 45(4)
Institute of Nuclear Physics, Polish Academy of Sciences (PL), University of Computer Sciences and Economics in Olsztyn (PL), University of Warmia and Mazury in Olsztyn (PL), Polish Academy of Sciences (PL)
Openalex Percentile: Top 15%
Radiation Dose and Imaging
1.36
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.