Strategies for the in-orbit gain tracking using the modulated X-ray sources for the Resolve microcalorimeter spectrometer on the X-ray Imaging and Spectroscopy Mission

Accurate and precise correction of the gain drift is the key to achieve the required energy resolution of the Resolve microcalorimeter spectrometer on the X-ray Imaging and Spectroscopy Mission (XRISM). Therefore, Resolve is equipped with highly configurable X-ray sources called the modulated X-ray source (MXS). The pulsed nature allows us to separate calibration and astrophysical X-rays by time interval selections. However, undesirable characteristics of the MXS, such as the afterglow X-rays, restrict the allowed configuration range. Moreover, the nonlinear and discontinuous behaviors of the calibration line count rate make the determination of the optimal setting highly complex. The MXS count rate model has been established using measurements in the spacecraft thermal vacuum test with the flight detector and MXS. A trade-off study using the model enables us to choose a few settings for the continuous use of the MXS optimized for different ranges of target gain tracking intervals. An alternative approach, where the MXS is used only intermittently, has also been developed and implemented. This new mode enables us to reconstruct the drift without having most of the undesirable effects in science data. This also forms the basis of the gain tracking under the current Resolve configuration with the closed gate valve.

Authors

Institutions

Publication Details

Journal
Journal of Astronomical Telescopes Instruments and Systems
Published
2025-10-16
DOI
https://doi.org/10.1117/1.jatis.11.4.042019
Primary Topic
Superconducting and THz Device Technology
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Strategies for the in-orbit gain tracking using the modulated X-ray sources for the Resolve microcalorimeter spectrometer on the X-ray Imaging and Spectroscopy Mission

Yoh Takei, R. F. Shipman, Ryuichi Fujimoto, Megan E. Eckart et al.
Journal of Astronomical Telescopes Instruments and Systems
Superconducting and THz Device Technology
article

Strategies for the in-orbit gain tracking using the modulated X-ray sources for the Resolve microcalorimeter spectrometer on the X-ray Imaging and Spectroscopy Mission

Yoh Takei, R. F. Shipman, Ryuichi Fujimoto, Megan E. Eckart, Renata Cumbee, Masahiro Tsujimoto, Maurice A. Leutenegger, F. S. Porter, Makoto Sawada, Yoshitaka Ishisaki, Shunji Kitamoto, Caroline A. Kilbourne, Cor de Vries
article en

Abstract

Accurate and precise correction of the gain drift is the key to achieve the required energy resolution of the Resolve microcalorimeter spectrometer on the X-ray Imaging and Spectroscopy Mission (XRISM). Therefore, Resolve is equipped with highly configurable X-ray sources called the modulated X-ray source (MXS). The pulsed nature allows us to separate calibration and astrophysical X-rays by time interval selections. However, undesirable characteristics of the MXS, such as the afterglow X-rays, restrict the allowed configuration range. Moreover, the nonlinear and discontinuous behaviors of the calibration line count rate make the determination of the optimal setting highly complex. The MXS count rate model has been established using measurements in the spacecraft thermal vacuum test with the flight detector and MXS. A trade-off study using the model enables us to choose a few settings for the continuous use of the MXS optimized for different ranges of target gain tracking intervals. An alternative approach, where the MXS is used only intermittently, has also been developed and implemented. This new mode enables us to reconstruct the drift without having most of the undesirable effects in science data. This also forms the basis of the gain tracking under the current Resolve configuration with the closed gate valve.

Journal of Astronomical Telescopes Instruments and SystemsVol. 11(04)
Lawrence Livermore National Laboratory (US), Goddard Space Flight Center (US), Rikkyo University (JP), University of Tsukuba (JP), National Sagamihara Hospital (JP), Space Research Organisation Netherlands (NL), Tokyo Metropolitan University (JP)
Openalex Percentile: Top 80%
Superconducting and THz Device Technology
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.