Assessment of Radiation Dose and Image Quality in Non-Contrast Head Computed Tomography: A Comparative Analysis Between 16-Slice and 32-Slice Multidetector Computed Tomography Scanners

Objectives Computed tomography (CT) has undergone rapid advancements over the past few decades, the main reason for which is the reduction of radiation dose and the improvement of image quality. We aimed to compare the radiation dose and image quality of 16-slice and 32-slice CT scanners using noise, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR). Material and Methods This retrospective study utilised data from 16- and 32-slice CT scanners from different imaging centers. The dose length product (DLP) value for CT plain brain was used to calculate the effective dose (ED) to compare the radiation dose values. Quantitative assessment of image quality was done using the region-of-interest (ROI) from distinct regions on the axial brain images. The image noise, SNR, and CNR were calculated and correlated with ED. Results The ED for the 16-slice CT was 1.46 mSv, while the ED for the 32-slice CT was 2.37 mSv. Moreover, higher quality images in terms of SNR and CNR were observed with the use of the 16-slice CT scanner compared to the 32-slice CT scanner. However, no correlation was observed between ED and noise. Conclusion This study concluded that the 16-slice CT scanner provides a safer choice for non-enhanced head CT with lower radiation exposure and produces improved image quality in terms of SNR and CNR than the 32-slice CT scanner; however, the use of reconstruction techniques like Adaptive Iterative Dose Reduction 3D (AIDR) reconstruction algorithm in the 16-slice CT scanner provides superior image quality and reduced radiation dose than the 32-slice CT scanner with the filtered back projection (FBP) technique.

Authors

Institutions

Publication Details

Journal
Journal of Health and Allied Sciences NU
Published
2025-10-16
DOI
https://doi.org/10.25259/jhs-2024-6-23-r1-(1442)
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Assessment of Radiation Dose and Image Quality in Non-Contrast Head Computed Tomography: A Comparative Analysis Between 16-Slice and 32-Slice Multidetector Computed Tomography Scanners

Manna Debnath, Santosh Ojha, Dolly A Sharma, M. Panda
Journal of Health and Allied Sciences NU
Radiation Dose and Imaging
article

Assessment of Radiation Dose and Image Quality in Non-Contrast Head Computed Tomography: A Comparative Analysis Between 16-Slice and 32-Slice Multidetector Computed Tomography Scanners

Manna Debnath, Santosh Ojha, Dolly A Sharma, M. Panda
article en

Abstract

Objectives Computed tomography (CT) has undergone rapid advancements over the past few decades, the main reason for which is the reduction of radiation dose and the improvement of image quality. We aimed to compare the radiation dose and image quality of 16-slice and 32-slice CT scanners using noise, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR). Material and Methods This retrospective study utilised data from 16- and 32-slice CT scanners from different imaging centers. The dose length product (DLP) value for CT plain brain was used to calculate the effective dose (ED) to compare the radiation dose values. Quantitative assessment of image quality was done using the region-of-interest (ROI) from distinct regions on the axial brain images. The image noise, SNR, and CNR were calculated and correlated with ED. Results The ED for the 16-slice CT was 1.46 mSv, while the ED for the 32-slice CT was 2.37 mSv. Moreover, higher quality images in terms of SNR and CNR were observed with the use of the 16-slice CT scanner compared to the 32-slice CT scanner. However, no correlation was observed between ED and noise. Conclusion This study concluded that the 16-slice CT scanner provides a safer choice for non-enhanced head CT with lower radiation exposure and produces improved image quality in terms of SNR and CNR than the 32-slice CT scanner; however, the use of reconstruction techniques like Adaptive Iterative Dose Reduction 3D (AIDR) reconstruction algorithm in the 16-slice CT scanner provides superior image quality and reduced radiation dose than the 32-slice CT scanner with the filtered back projection (FBP) technique.

Journal of Health and Allied Sciences NUVol. 16
Rayat Bahra University (IN), Shree Guru Gobind Singh Tricentenary University (IN), Charotar University of Science and Technology (IN), Environmental and Occupational Health Sciences Institute (US)
Openalex Percentile: Top 57%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.