Ultrasound 2D-SWE inter-system variability and reliability: Insights from an elasticity phantom model

Two-dimensional shear wave elastography (2D-SWE) enables quantitative assessment of tissue stiffness. However, due to system-specific variability in shear wave generation, cross-platform reproducibility needs to be assessed. Therefore, this study aimed to assess the inter-system agreement of 2D-SWE measurements between two widely used ultrasound systems (Philips EPIQ and GE LOGIQ) using a standardized tissue-mimicking elasticity phantom with known stiffness values. A total of 320 elasticity measurements, expressed in kilopascals (kPa), were obtained from four lesions representing varying stiffness levels. A total of 160 measurements were obtained per system, comprising 40 measurements for each lesion. Inter-system agreement of 2D-SWE measurements was asssessed using the intraclass correlation coefficient (ICC) and Bland-Altman plot, and the differences in 2D-SWE measurements between the two ultrasound systems were assessed using the Wilcoxon-rank test. Significant differences were observed in stiffness measurements. For Lesion I (reference: 8 ± 3 kPa), the EPIQ overestimated stiffness (9.72 ± 0.93 kPa), while the LOGIQ closely aligned with the reference (7.95 ± 0.44 kPa). Both systems underestimated high stiffness levels, such as Lesion IV (reference: 80 ± 8 kPa), reporting 11.23 ± 2.04 kPa (EPIQ) and 20.33 ± 2.64 kPa (LOGIQ). The ICC for inter-system agreement was 0.81 (p < 0.0001), indicating good overall consistency. Bias in measurements was 0.78 ± 6.45 kPa. There were significant differences in the 2D-SWE measurements for all lesions between the two ultrasound systems (p < 0.0001). These findings highlight the need for standardization and calibration protocols to address cross-platform inconsistencies. Future research should focus on validating findings in clinical settings and exploring solutions to improve reproducibility and reliability across systems.

Authors

Institutions

Publication Details

Journal
Journal of Radiation Research and Applied Sciences
Published
2025-10-16
DOI
https://doi.org/10.1016/j.jrras.2025.102022
Primary Topic
Ultrasound Imaging and Elastography
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Ultrasound 2D-SWE inter-system variability and reliability: Insights from an elasticity phantom model

Salahaden R. Sultan
Journal of Radiation Research and Applied Sciences
Ultrasound Imaging and Elastography
article

Ultrasound 2D-SWE inter-system variability and reliability: Insights from an elasticity phantom model

Salahaden R. Sultan
article en

Abstract

Two-dimensional shear wave elastography (2D-SWE) enables quantitative assessment of tissue stiffness. However, due to system-specific variability in shear wave generation, cross-platform reproducibility needs to be assessed. Therefore, this study aimed to assess the inter-system agreement of 2D-SWE measurements between two widely used ultrasound systems (Philips EPIQ and GE LOGIQ) using a standardized tissue-mimicking elasticity phantom with known stiffness values. A total of 320 elasticity measurements, expressed in kilopascals (kPa), were obtained from four lesions representing varying stiffness levels. A total of 160 measurements were obtained per system, comprising 40 measurements for each lesion. Inter-system agreement of 2D-SWE measurements was asssessed using the intraclass correlation coefficient (ICC) and Bland-Altman plot, and the differences in 2D-SWE measurements between the two ultrasound systems were assessed using the Wilcoxon-rank test. Significant differences were observed in stiffness measurements. For Lesion I (reference: 8 ± 3 kPa), the EPIQ overestimated stiffness (9.72 ± 0.93 kPa), while the LOGIQ closely aligned with the reference (7.95 ± 0.44 kPa). Both systems underestimated high stiffness levels, such as Lesion IV (reference: 80 ± 8 kPa), reporting 11.23 ± 2.04 kPa (EPIQ) and 20.33 ± 2.64 kPa (LOGIQ). The ICC for inter-system agreement was 0.81 (p < 0.0001), indicating good overall consistency. Bias in measurements was 0.78 ± 6.45 kPa. There were significant differences in the 2D-SWE measurements for all lesions between the two ultrasound systems (p < 0.0001). These findings highlight the need for standardization and calibration protocols to address cross-platform inconsistencies. Future research should focus on validating findings in clinical settings and exploring solutions to improve reproducibility and reliability across systems.

Journal of Radiation Research and Applied SciencesVol. 18(4)
King Abdulaziz University (SA)
Openalex Percentile: Top 57%
Ultrasound Imaging and Elastography
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.