Benchmarking of the efficiency of the NaI(Tl) detector using FLUKA, MCNP, and experimental measurements at multiple source distances
Accurate calibration of NaI(Tl) scintillation detectors is essential for reliable gamma-ray spectrometry in nuclear safeguards and environmental radiation monitoring. This study investigates the full energy peak efficiency (FEPE) of a NaI(Tl) detector using experimental measurements and Monte Carlo simulation performed with FLUKA and MCNP. Standard point sources ( , , and ) were measured at four source-detector distances (4, 8, 12, and 16 cm) covering photon energies from 356 to 1333 keV. Both simulation tools demonstrated an average deviation of less then 6 % and a maximum deviation of 18 % from the experimental results in all configurations. Furthermore, source misalignment was numerically evaluated and found to have a measurable impact only at short source-detector distance. This work represents the first integrated benchmarking of FLUKA and MCNP against experimental FEPE data across multiple geometries, offering a validated framework for accurate detector modeling and calibration. • Monte Carlo-based FEPE benchmarking using FLUKA and MCNP. • Evaluation of geometric efficiency changes under misalignment. • Validation with experimental measurements for reliability.
Authors
- A. Jalil (ORCID: https://orcid.org/0000-0003-2645-1494)
- Javed Hussain (ORCID: https://orcid.org/0009-0006-6807-4219)
- Ghulam Hussan
- Da Chen (ORCID: https://orcid.org/0000-0002-3125-4033)
- M. Shoaib (ORCID: https://orcid.org/0000-0001-6791-8252)
- Zeenat Ullah
- Xin Xianpeng
- Sang Yingming
Institutions
- Islamia University of Bahawalpur (PK)
- Kohat University of Science and Technology (PK)
- National Centre for Nuclear Energy, Science and Technology (MA)
- Abdus Salam Centre for Physics (PK)
- Southeast University (CN)
Publication Details
- Journal
- Nuclear Analysis
- Published
- 2025-10-15
- DOI
- https://doi.org/10.1016/j.nucana.2025.100185
- Citations
- 1
- Primary Topic
- Radiation Detection and Scintillator Technologies
- Type
- article
- Field-Weighted Citation Impact
- 1.88