Effect of intraoperative delayed time on correction accuracy in 1050 Hz excimer laser-assisted in situ keratomileusis

This study aimed to analyze and compare the effects of intraoperative delayed time (IDT) on refractive correction accuracy in femtosecond laser-assisted stromal in situ keratomileusis (FS-LASIK) using 1050 Hz excimer laser equipment. Eighty-one eyes of 81 myopic and astigmatic patients were divided into three groups according to the IDT: mild delay (IDT < 5 s, 29 eyes), moderate delay (5 s ≤ IDT < 10 s, 26 eyes), and severe delay (10 s ≤ IDT ≤ 60 s, 26 eyes) groups. Uncorrected distance visual acuity (UDVA), spherical equivalent (SE), and intraocular pressure (IOP) were measured 1 week, 1 month, and 3 months postoperatively. The static cyclotorsion and dynamic cyclotorsion components (SCC and DCC, respectively) were recorded. Central corneal thickness was measured before and 1 week after the operation. Differences in the UDVA, refractive power correction error (RPCE), and corneal ablation error (CAE) between the different groups were compared. The postoperative UDVA of the three groups was good; no significant difference as observed in the SCC and RPCE 1 week after surgery. The DCC and RPCE at 1 and 3 months postoperatively in the three groups significantly varied. For patients with an IDT less than 10 s, no significant difference was observed between the actual and theoretical ablation corneal thicknesses. However, for those with IDT longer than 10 s, the actual ablation thickness was greater than the theoretical ablation thickness. The CAE increased with the IDT lengthening, and the average actual ablation depth increased by 0.09 μm for every extra second. Education and training on preoperative ocular fixation should be emphasized. Patient delay in laser ablation due to poor cooperation may still lead to RPCE and CAE, even though the 1050 Hz excimer laser has the highest ablation speed, highest tracking speed, and seven-dimensional eye movement tracking.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2025-10-15
DOI
https://doi.org/10.1038/s41598-025-15742-x
Primary Topic
Corneal surgery and disorders
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Effect of intraoperative delayed time on correction accuracy in 1050 Hz excimer laser-assisted in situ keratomileusis

Shaozhen Zhao, Wenjie Qi, Fei Ran Li, Guiqin Wang et al.
Scientific Reports
Corneal surgery and disorders
article

Effect of intraoperative delayed time on correction accuracy in 1050 Hz excimer laser-assisted in situ keratomileusis

Shaozhen Zhao, Wenjie Qi, Fei Ran Li, Guiqin Wang, Yang Liu, Ruibo Yang
article en

Abstract

This study aimed to analyze and compare the effects of intraoperative delayed time (IDT) on refractive correction accuracy in femtosecond laser-assisted stromal in situ keratomileusis (FS-LASIK) using 1050 Hz excimer laser equipment. Eighty-one eyes of 81 myopic and astigmatic patients were divided into three groups according to the IDT: mild delay (IDT < 5 s, 29 eyes), moderate delay (5 s ≤ IDT < 10 s, 26 eyes), and severe delay (10 s ≤ IDT ≤ 60 s, 26 eyes) groups. Uncorrected distance visual acuity (UDVA), spherical equivalent (SE), and intraocular pressure (IOP) were measured 1 week, 1 month, and 3 months postoperatively. The static cyclotorsion and dynamic cyclotorsion components (SCC and DCC, respectively) were recorded. Central corneal thickness was measured before and 1 week after the operation. Differences in the UDVA, refractive power correction error (RPCE), and corneal ablation error (CAE) between the different groups were compared. The postoperative UDVA of the three groups was good; no significant difference as observed in the SCC and RPCE 1 week after surgery. The DCC and RPCE at 1 and 3 months postoperatively in the three groups significantly varied. For patients with an IDT less than 10 s, no significant difference was observed between the actual and theoretical ablation corneal thicknesses. However, for those with IDT longer than 10 s, the actual ablation thickness was greater than the theoretical ablation thickness. The CAE increased with the IDT lengthening, and the average actual ablation depth increased by 0.09 μm for every extra second. Education and training on preoperative ocular fixation should be emphasized. Patient delay in laser ablation due to poor cooperation may still lead to RPCE and CAE, even though the 1050 Hz excimer laser has the highest ablation speed, highest tracking speed, and seven-dimensional eye movement tracking.

Scientific ReportsVol. 15(1)
Tianjin Medical University Eye Hospital (CN), Tianjin Medical University (CN)
Openalex Percentile: Top 57%
Corneal surgery and disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.