Clinical Assessment of a Virtual Reality Perimeter Versus the Humphrey Field Analyzer: Comparative Reliability, Usability, and Prospective Applications

<i>Background:</i> This study compared the performance of a Head-mounted Virtual Reality Perimeter (HVRP) with the Humphrey Field Analyzer (HFA), the standard in automated perimetry. The HFA is the established standard for automated perimetry but is constrained by lengthy testing, bulky equipment, and limited patient comfort. Comparative data on newer head-mounted virtual reality perimeters are limited, leaving uncertainty about their clinical reliability and potential advantages. <i>Aim:</i> The aim was to evaluate parameters such as visual field outcomes, portability, patient comfort, eye tracking, and usability. <i>Methods:</i> Participants underwent testing with both devices, assessing metrics like mean deviation (MD), pattern standard deviation (PSD), and duration. <i>Results:</i> The HVRP demonstrated small but statistically significant differences in MD and PSD compared to the HFA, while maintaining a consistent trend across participants. MD values were slightly more negative for HFA than HVRP (average difference -0.60 dB, <i>p</i> = 0.0006), while pattern standard deviation was marginally higher with HFA (average difference 0.38 dB, <i>p</i> = 0.00018). Although statistically significant, these differences were small in magnitude and do not undermine the clinical utility or reproducibility of the device. Notably, HVRP showed markedly shorter testing times with HVRP (7.15 vs. 18.11 min, mean difference 10.96 min, <i>p</i> < 0.0001). Its lightweight, portable design allowed for bedside and home testing, enhancing accessibility for pediatric, geriatric, and mobility-impaired patients. Participants reported greater comfort due to the headset design, which eliminated the need for chin rests. The device also offers potential for AI integration and remote data analysis. <i>Conclusions:</i> The HVRP proved to be a reliable, user-friendly alternative to traditional perimetry. Its advantages in comfort, portability, and test efficiency support its use in both clinical settings and remote screening programs for visual field assessment. Its portability and user-friendly design support broader use in clinical practice and expand possibilities for bedside assessment, home monitoring, and remote screening, particularly in populations with limited access to conventional perimetry.

Authors

Institutions

Publication Details

Journal
Vision
Published
2025-10-11
DOI
https://doi.org/10.3390/vision9040086
Citations
2
Primary Topic
Medical Imaging and Analysis
Type
article
Field-Weighted Citation Impact
1.35
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Clinical Assessment of a Virtual Reality Perimeter Versus the Humphrey Field Analyzer: Comparative Reliability, Usability, and Prospective Applications

Francesco Pellegrini, F. Cappellani, Marco Zeppieri, Caterina Gagliano et al.
2 citations
Vision
Medical Imaging and Analysis
1.35
article

Clinical Assessment of a Virtual Reality Perimeter Versus the Humphrey Field Analyzer: Comparative Reliability, Usability, and Prospective Applications

Francesco Pellegrini, F. Cappellani, Marco Zeppieri, Caterina Gagliano, Alessandra Cuna, Fabiana D’Esposito, Roberta Amato, Alessandro Avitabile, F Visalli
article en
2 citations

Abstract

<i>Background:</i> This study compared the performance of a Head-mounted Virtual Reality Perimeter (HVRP) with the Humphrey Field Analyzer (HFA), the standard in automated perimetry. The HFA is the established standard for automated perimetry but is constrained by lengthy testing, bulky equipment, and limited patient comfort. Comparative data on newer head-mounted virtual reality perimeters are limited, leaving uncertainty about their clinical reliability and potential advantages. <i>Aim:</i> The aim was to evaluate parameters such as visual field outcomes, portability, patient comfort, eye tracking, and usability. <i>Methods:</i> Participants underwent testing with both devices, assessing metrics like mean deviation (MD), pattern standard deviation (PSD), and duration. <i>Results:</i> The HVRP demonstrated small but statistically significant differences in MD and PSD compared to the HFA, while maintaining a consistent trend across participants. MD values were slightly more negative for HFA than HVRP (average difference -0.60 dB, <i>p</i> = 0.0006), while pattern standard deviation was marginally higher with HFA (average difference 0.38 dB, <i>p</i> = 0.00018). Although statistically significant, these differences were small in magnitude and do not undermine the clinical utility or reproducibility of the device. Notably, HVRP showed markedly shorter testing times with HVRP (7.15 vs. 18.11 min, mean difference 10.96 min, <i>p</i> < 0.0001). Its lightweight, portable design allowed for bedside and home testing, enhancing accessibility for pediatric, geriatric, and mobility-impaired patients. Participants reported greater comfort due to the headset design, which eliminated the need for chin rests. The device also offers potential for AI integration and remote data analysis. <i>Conclusions:</i> The HVRP proved to be a reliable, user-friendly alternative to traditional perimetry. Its advantages in comfort, portability, and test efficiency support its use in both clinical settings and remote screening programs for visual field assessment. Its portability and user-friendly design support broader use in clinical practice and expand possibilities for bedside assessment, home monitoring, and remote screening, particularly in populations with limited access to conventional perimetry.

VisionVol. 9(4)
University of Trieste (IT), Università degli Studi di Enna Kore (IT), University of Catania (IT), AULSS 2 Marca Trevigiana (IT), Azienda Ospedaliera Santa Maria Degli Angeli (IT), Eye Center (SA), Imperial College London (GB)
Openalex Percentile: Top 17%
Medical Imaging and Analysis
1.35
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.