Design of a PET/MRI DOI Detector Consisting of a Quasi-Block Scintillator and a Dual-Sided Readout Photosensor with MLPE in a High Magnetic Field

The spatial resolution of small animal positron emission tomography/magnetic resonance imaging (PET/MRI) detectors for operating in a high magnetic field, which use very small and thin scintillation pixels, is degraded at the periphery of the field of view (FOV). To solve this problem, we designed a detector that measures the depth of interaction (DOI) when gamma rays, a form of electromagnetic radiation, interact with scintillation pixels. The detector consists of four layers of quasi-block scintillators with photosensors with no signal distortion in a magnetic field at both ends of each layer, and the location of scintillation events was determined using a lookup table and maximum likelihood position estimation (MLPE). DETECT2000 simulation was performed to evaluate the performance of the designed detector. A lookup table was created using the light signals collected from the photosensors, and the location was determined by comparing it with new signals with reference to the MLPE. As a result, the discrimination accuracy was excellent at 92.0%. The application of this detector to small animal PET/MRI is expected to deliver outstanding spatial resolution.

Authors

Publication Details

Journal
Journal of Magnetics
Published
2025-09-30
DOI
https://doi.org/10.4283/jmag.2025.30.3.448
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Design of a PET/MRI DOI Detector Consisting of a Quasi-Block Scintillator and a Dual-Sided Readout Photosensor with MLPE in a High Magnetic Field

Cheol-Ha Baek, Seung‐Jae Lee
Journal of Magnetics
Medical Imaging Techniques and Applications
article

Design of a PET/MRI DOI Detector Consisting of a Quasi-Block Scintillator and a Dual-Sided Readout Photosensor with MLPE in a High Magnetic Field

Cheol-Ha Baek, Seung‐Jae Lee
article en

Abstract

The spatial resolution of small animal positron emission tomography/magnetic resonance imaging (PET/MRI) detectors for operating in a high magnetic field, which use very small and thin scintillation pixels, is degraded at the periphery of the field of view (FOV). To solve this problem, we designed a detector that measures the depth of interaction (DOI) when gamma rays, a form of electromagnetic radiation, interact with scintillation pixels. The detector consists of four layers of quasi-block scintillators with photosensors with no signal distortion in a magnetic field at both ends of each layer, and the location of scintillation events was determined using a lookup table and maximum likelihood position estimation (MLPE). DETECT2000 simulation was performed to evaluate the performance of the designed detector. A lookup table was created using the light signals collected from the photosensors, and the location was determined by comparing it with new signals with reference to the MLPE. As a result, the discrimination accuracy was excellent at 92.0%. The application of this detector to small animal PET/MRI is expected to deliver outstanding spatial resolution.

Journal of MagneticsVol. 30(3)
Openalex Percentile: Top 57%
Medical Imaging Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.