Achieving high-resolution spatial mapping of dose and linear energy transfer in ion-beam therapy

<i>Objective.</i>We present a novel detector system capable of simultaneously obtaining two-dimensional (2D) dose and linear energy transfer (LET) maps from Al<sub>2</sub>O<sub>3</sub>films of sizes up to(190×190) mm<sup>2</sup>using optically stimulated luminescence (OSL).<i>Approach.</i>The system is based on the OSL from doped Al<sub>2</sub>O<sub>3</sub>films measured using pulsed stimulation and time-resolved OSL detection. The system was calibrated in<sup>1</sup>H-,<sup>4</sup>He-, and<sup>12</sup>C-ion beams. The minimal detectable dose (MDD), response homogeneity, pixel cross-stimulation, and precision were also investigated. To demonstrate the system's performance, 2D maps of dose and LET of a proton Bragg curve were measured.<i>Main results.</i>The system was demonstrated to be capable of obtaining absorbed dose and LET 2D maps with a spatial resolution set to(1.0×1.0) mm<sup>2</sup>. The system was calibrated over a clinically relevant range of LET values and characterized through assessment of MDD (<25 mGy), readout protocols, response homogeneity, precision, and relative detector efficiency. Dose and LET were found to be measurable with precisions of 3.2% and 10%, respectively.<i>Significance.</i>To our knowledge, this is the first passive detector system capable of measuring LET and dose simultaneously in 2D with a 1.0 mm<sup>2</sup>spatial resolution requiring no prior information. Such properties are required to experimentally validate simulated dose and LET distributions in ion-beam therapy, paving the way for sub-mm mapping of radiation-induced relative biological effectiveness.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2025-10-13
DOI
https://doi.org/10.1088/1361-6560/ae128f
Primary Topic
Radiation Detection and Scintillator Technologies
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Achieving high-resolution spatial mapping of dose and linear energy transfer in ion-beam therapy

Jeppe Brage Christensen, Stephan Brons, E.G. Yukihara, Iván Domingo Muñoz et al.
Physics in Medicine and Biology
Radiation Detection and Scintillator Technologies
article

Achieving high-resolution spatial mapping of dose and linear energy transfer in ion-beam therapy

Jeppe Brage Christensen, Stephan Brons, E.G. Yukihara, Iván Domingo Muñoz, Michele Togno, Mads L. Jensen, O. Jaekel
article en

Abstract

<i>Objective.</i>We present a novel detector system capable of simultaneously obtaining two-dimensional (2D) dose and linear energy transfer (LET) maps from Al<sub>2</sub>O<sub>3</sub>films of sizes up to(190×190) mm<sup>2</sup>using optically stimulated luminescence (OSL).<i>Approach.</i>The system is based on the OSL from doped Al<sub>2</sub>O<sub>3</sub>films measured using pulsed stimulation and time-resolved OSL detection. The system was calibrated in<sup>1</sup>H-,<sup>4</sup>He-, and<sup>12</sup>C-ion beams. The minimal detectable dose (MDD), response homogeneity, pixel cross-stimulation, and precision were also investigated. To demonstrate the system's performance, 2D maps of dose and LET of a proton Bragg curve were measured.<i>Main results.</i>The system was demonstrated to be capable of obtaining absorbed dose and LET 2D maps with a spatial resolution set to(1.0×1.0) mm<sup>2</sup>. The system was calibrated over a clinically relevant range of LET values and characterized through assessment of MDD (<25 mGy), readout protocols, response homogeneity, precision, and relative detector efficiency. Dose and LET were found to be measurable with precisions of 3.2% and 10%, respectively.<i>Significance.</i>To our knowledge, this is the first passive detector system capable of measuring LET and dose simultaneously in 2D with a 1.0 mm<sup>2</sup>spatial resolution requiring no prior information. Such properties are required to experimentally validate simulated dose and LET distributions in ion-beam therapy, paving the way for sub-mm mapping of radiation-induced relative biological effectiveness.

Physics in Medicine and BiologyVol. 70(21)
German Cancer Research Center (DE), Heidelberg University (DE), University Hospital Heidelberg (DE), Paul Scherrer Institute (CH)
National Science Foundation, Schweizerischer Nationalfonds zur Förderung der Wissenschaftlichen Forschung, Universitätsspital Zürich
Openalex Percentile: Top 61%
Radiation Detection and Scintillator Technologies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.