Development of STM8 and Kalman Filter Based Heater Controller for Automatic Tissue Processor

A tissue processor is a device that automatically processes tissue into several steps for histological studies in medical or biological laboratories. One of the stages of this device is embedding, where the tissue is inserted into a vessel that is heated to a stable temperature of 60 to 80°C so that the paraffin remains liquid during this process. The handling of tissue specimens is a crucial aspect of ensuring an accurate diagnosis of the tissue. However, due to its high cost, it can affect scientific development that requires tissue processing. Reverse engineering is one of the methods commonly used to develop a tool at a lower price; one way is to change the essential components used while still maintaining the device's primary function. This research aims to develop a heater controller for a tissue processor using STM8 and LM35 analog sensors equipped with a Kalman filter. The design of the Kalman filter will be determined based on its component equations. The measurement and variance constant values will be simulated and analysed in MATLAB for the most suitable parameter value. Afterwards, the proposed design will be conducted using the Arduino Integrated Development Environment (IDE) to reduce the noise in the LM35 sensor reading and then saved using ArduSpreadsheet. The simulation and implementation results provide evidence that the proposed Kalman filter effectively filters signals contaminated with noise, even in the presence of high data variations that may occur in sensor readings. In testing, it is necessary to pay attention to the ratio that can reduce noise while keeping the characteristics of the sensor itself. Numerous investigations have explored various ratios, recommending an appropriate ratio of 100. Based on the implementation of this ratio with measurement constant (R) 10 and variance constant (Q) 0.1, resulted in a 0.29°C mean error and stable sensor readings.

Authors

Institutions

Publication Details

Journal
Engineering headway
Published
2025-10-13
DOI
https://doi.org/10.4028/p-1iwrkp
Primary Topic
Fault Detection and Control Systems
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development of STM8 and Kalman Filter Based Heater Controller for Automatic Tissue Processor

Jans Hendry, Hidayat Nur Isnianto, Naufal Rashad Aryaputra, Ahmad Nugraha Bayu Mukti
Engineering headway
Fault Detection and Control Systems
article

Development of STM8 and Kalman Filter Based Heater Controller for Automatic Tissue Processor

Jans Hendry, Hidayat Nur Isnianto, Naufal Rashad Aryaputra, Ahmad Nugraha Bayu Mukti
article en

Abstract

A tissue processor is a device that automatically processes tissue into several steps for histological studies in medical or biological laboratories. One of the stages of this device is embedding, where the tissue is inserted into a vessel that is heated to a stable temperature of 60 to 80°C so that the paraffin remains liquid during this process. The handling of tissue specimens is a crucial aspect of ensuring an accurate diagnosis of the tissue. However, due to its high cost, it can affect scientific development that requires tissue processing. Reverse engineering is one of the methods commonly used to develop a tool at a lower price; one way is to change the essential components used while still maintaining the device's primary function. This research aims to develop a heater controller for a tissue processor using STM8 and LM35 analog sensors equipped with a Kalman filter. The design of the Kalman filter will be determined based on its component equations. The measurement and variance constant values will be simulated and analysed in MATLAB for the most suitable parameter value. Afterwards, the proposed design will be conducted using the Arduino Integrated Development Environment (IDE) to reduce the noise in the LM35 sensor reading and then saved using ArduSpreadsheet. The simulation and implementation results provide evidence that the proposed Kalman filter effectively filters signals contaminated with noise, even in the presence of high data variations that may occur in sensor readings. In testing, it is necessary to pay attention to the ratio that can reduce noise while keeping the characteristics of the sensor itself. Numerous investigations have explored various ratios, recommending an appropriate ratio of 100. Based on the implementation of this ratio with measurement constant (R) 10 and variance constant (Q) 0.1, resulted in a 0.29°C mean error and stable sensor readings.

Engineering headwayVol. 27
Universitas Gadjah Mada (ID), Micos Engineering (Switzerland) (CH)
Openalex Percentile: Top 63%
Fault Detection and Control Systems
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.