Quantitative terbium-161 SPECT/CT imaging: demonstrating the feasibility of image-based dosimetry and highlighting pitfalls
This study demonstrates that highly accurate SPECT/CT-based <sup>161</sup>Tb quantification is feasible, further cementing <sup>161</sup>Tb as a viable theragnostic alternative. A MEGP collimator, a 75 keV window, and a CF derived from a homogeneous cylinder measurement appears preferable. The 49 keV window could be useful at late imaging time points, given its high sensitivity, if further optimized. Degradation from penetration and subsequent downscatter may be mitigated with a more refined reconstruction. Further investigations into dead-time effects are encouraged.