Carbon fibre-enforced stabilization system for ministernotomy resynthesis
Sternal dehiscence incidence ranges from 0.2% to 5%]. Risk factors include obesity, advanced age, use of bilateral internal thoracic arteries, diabetes mellitus and chronic obstructive pulmonary disease [1]. Traditional techniques with steel wires have been associated with dehiscence rates of up to 11.8% [2]. Utilizing titanium plates, the incidence of sternal dehiscence has reduced to 1.5% [3]. The STERN FIX Sternal Stabilization System is a biocompatible carbon-fibre reinforced poly-ether-ether-ketone tool constituting two components, both with a curved arm that embraces the sternum at the level of an intercostal space and does not perforate the intercostal fascia, reducing the retrosternal bleeding [4]. The sternal closure is performed following five steps: sizing using a gauger (device available for sternal bone thicknesses between 9.5 and 17mm); marking the selected intercostal space bilaterally to allow the passage of the device; placement of both device parts; closure of the device; and cutting the excess segment. The sternal closure is complemented with cerclage wire. The first results are very satisfactory, with 0% sternal dehiscence at 6 months on 30 patients [5], comparable to those achieved by adopting the Robicsek technique [6]. We present a video tutorial case of ministernotomy closure using the STERN FIX system.
Authors
- Gustavo Woll
- Lorena Mujica
- Manuel Castellà (ORCID: https://orcid.org/0000-0001-9274-2121)
- Bruno Chiarello (ORCID: https://orcid.org/0000-0002-6223-1619)
- Sherif Negm (ORCID: https://orcid.org/0000-0001-7827-8334)
- Brayan Rubio
Institutions
- Universitat de Barcelona (ES)
Publication Details
- Journal
- Multimedia Manual of Cardio-Thoracic Surgery
- Published
- 2025-10-13
- DOI
- https://doi.org/10.1510/mmcts.2025.058
- Primary Topic
- Reconstructive Surgery and Microvascular Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00