A highly sensitive headspace gas chromatographic method fully optimized for fast routine determination of carboxyhemoglobin in postmortem blood

In some cases, the determination of carboxyhemoglobin (COHb) concentration in postmortem blood using standard spectrophotometric methods, may be either impossible or yield misleading results. In forensic laboratories, there is an increasing need to analyze COHb saturation in autopsy blood samples, the quality of which is often compromised by various factors. Therefore, a gas chromatographic method has been developed and validated to confirm the results of an established spectrophotometric method and to determine the concentration of COHb in postmortem samples affected by processes such as putrefaction or thermocoagulation. This method is based on automated headspace analysis of a blood sample treated with a solution of potassium hexacyanoferrate and saponin. It employs a combination of two capillary columns for chromatographic separation, followed by a highly sensitive detection system that incorporates a methanizer to convert carbon monoxide (CO) to methane couple to a flame ionization detector (FID). The percentage of COHb concentration is calculated using the ratio of the analyzed blood to that of fully CO-saturated blood. Careful optimization of the method has reduced the sample requirement to 100 mg while maintaining high sensitivity. Additionally, the sample pretreatment process has been simplified to ensure that the method can be easily integrated into the routine operation of a conventional forensic laboratory. The method demonstrates linearity across a concentration range of 0.1% to 100% COHb. The determined values of LOD (< 0.01% COHb) and LOQ (0.1% COHb) reflect excellent sensitivity. Overall, the validation study confirmed the method for its intended purpose. The method is particularly effective in reliably determining COHb content in putrefied and similarly degraded samples. Its robustness has been further validated through the analysis of dozens of real samples over more than four years.

Authors

Institutions

Publication Details

Journal
Journal of Analytical Toxicology
Published
2025-10-08
DOI
https://doi.org/10.1093/jat/bkaf089
Primary Topic
Metabolomics and Mass Spectrometry Studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A highly sensitive headspace gas chromatographic method fully optimized for fast routine determination of carboxyhemoglobin in postmortem blood

Lucie Borovcová, Petr Handlos, Vladimíra Gebauerová, Roman Papoušek et al.
Journal of Analytical Toxicology
Metabolomics and Mass Spectrometry Studies
article

A highly sensitive headspace gas chromatographic method fully optimized for fast routine determination of carboxyhemoglobin in postmortem blood

Lucie Borovcová, Petr Handlos, Vladimíra Gebauerová, Roman Papoušek, M Staňková, Edita Červenková, Denisa Ptáčková
article en

Abstract

In some cases, the determination of carboxyhemoglobin (COHb) concentration in postmortem blood using standard spectrophotometric methods, may be either impossible or yield misleading results. In forensic laboratories, there is an increasing need to analyze COHb saturation in autopsy blood samples, the quality of which is often compromised by various factors. Therefore, a gas chromatographic method has been developed and validated to confirm the results of an established spectrophotometric method and to determine the concentration of COHb in postmortem samples affected by processes such as putrefaction or thermocoagulation. This method is based on automated headspace analysis of a blood sample treated with a solution of potassium hexacyanoferrate and saponin. It employs a combination of two capillary columns for chromatographic separation, followed by a highly sensitive detection system that incorporates a methanizer to convert carbon monoxide (CO) to methane couple to a flame ionization detector (FID). The percentage of COHb concentration is calculated using the ratio of the analyzed blood to that of fully CO-saturated blood. Careful optimization of the method has reduced the sample requirement to 100 mg while maintaining high sensitivity. Additionally, the sample pretreatment process has been simplified to ensure that the method can be easily integrated into the routine operation of a conventional forensic laboratory. The method demonstrates linearity across a concentration range of 0.1% to 100% COHb. The determined values of LOD (< 0.01% COHb) and LOQ (0.1% COHb) reflect excellent sensitivity. Overall, the validation study confirmed the method for its intended purpose. The method is particularly effective in reliably determining COHb content in putrefied and similarly degraded samples. Its robustness has been further validated through the analysis of dozens of real samples over more than four years.

Journal of Analytical Toxicology
University of Pardubice (CZ), University of Ostrava (CZ), University Hospital Ostrava (CZ), Palacký University Olomouc (CZ)
Openalex Percentile: Top 75%
Metabolomics and Mass Spectrometry Studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.