Impact of hemolysis on coagulation testing: comparative analysis of two different analyzers using photometric and mechanical clot detection techniques

BACKGROUND: Analytical methods for clot detection may differ in susceptibility to hemolysis, especially photometric versus mechanical systems.METHODS: Two coagulation analyzers have been compared: Werfen ACL TOP 750, photometric clot detection versus Stago sthemO 301, mechanical clot detection, using 100 routine plasma samples (50 hemolyzed, 50 non-hemolyzed). Prothrombin time (PT), activated partial thromboplastin time (APTT) and fibrinogen were measured in duplicate per analyzer. Analytical imprecision (coefficient of variation; CV%), correlation (Spearman’s rank) and linear fit were calculated, while differences were evaluated using the Mann-Whitney U-Test. Bootstrap and hypothesis tests for equality of regression slopes and intercepts were used to compare inter-analyzer relationships in non-hemolyzed and hemolyzed plasma samples.RESULTS: PT, APTT and fibrinogen values did not significantly differ between non-hemolyzed and hemolyzed samples (all p>0.05). Correlation between analyzers was strong across all parameters (all r>0.74). Bootstrap test revealed no significant differences of inter-analyzer correlations between non-hemolyzed and hemolyzed samples for all parameters (all p>0.05). The hypothesis tests for equality of regression slopes and intercepts showed significant differences between non-hemolyzed and hemolyzed samples for fibrinogen but not for PT and APTT. Imprecision was higher with ACL TOP, especially for APTT in hemolyzed samples (CV% 1.6 versus 0.6%, p<0.001). Fibrinogen measurements on ACL TOP displayed approximately double CV% compared to Stago sthemO, irrespective of hemolysis.CONCLUSIONS: Both analyzers showed comparable performance for PT and APTT in hemolyzed and non-hemolyzed samples, while fibrinogen displayed a different trend. Mechanical clot detection provided better overall precision for fibrinogen and for APTT in hemolyzed specimens.

Authors

Institutions

Publication Details

Journal
Biochimica Clinica
Published
2025-10-10
DOI
https://doi.org/10.23736/s0393-0564.25.00096-2
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Impact of hemolysis on coagulation testing: comparative analysis of two different analyzers using photometric and mechanical clot detection techniques

Simone De Nitto, Emmanuel J. Favaloro, Davide Demonte, Giuseppe Lippi et al.
Biochimica Clinica
Clinical Laboratory Practices and Quality Control
article

Impact of hemolysis on coagulation testing: comparative analysis of two different analyzers using photometric and mechanical clot detection techniques

Simone De Nitto, Emmanuel J. Favaloro, Davide Demonte, Giuseppe Lippi, Gian Luca Salvagno, Elia Ponchini, Nicola Baratto, Alessandro Lorenzetto
article en

Abstract

BACKGROUND: Analytical methods for clot detection may differ in susceptibility to hemolysis, especially photometric versus mechanical systems.METHODS: Two coagulation analyzers have been compared: Werfen ACL TOP 750, photometric clot detection versus Stago sthemO 301, mechanical clot detection, using 100 routine plasma samples (50 hemolyzed, 50 non-hemolyzed). Prothrombin time (PT), activated partial thromboplastin time (APTT) and fibrinogen were measured in duplicate per analyzer. Analytical imprecision (coefficient of variation; CV%), correlation (Spearman’s rank) and linear fit were calculated, while differences were evaluated using the Mann-Whitney U-Test. Bootstrap and hypothesis tests for equality of regression slopes and intercepts were used to compare inter-analyzer relationships in non-hemolyzed and hemolyzed plasma samples.RESULTS: PT, APTT and fibrinogen values did not significantly differ between non-hemolyzed and hemolyzed samples (all p>0.05). Correlation between analyzers was strong across all parameters (all r>0.74). Bootstrap test revealed no significant differences of inter-analyzer correlations between non-hemolyzed and hemolyzed samples for all parameters (all p>0.05). The hypothesis tests for equality of regression slopes and intercepts showed significant differences between non-hemolyzed and hemolyzed samples for fibrinogen but not for PT and APTT. Imprecision was higher with ACL TOP, especially for APTT in hemolyzed samples (CV% 1.6 versus 0.6%, p<0.001). Fibrinogen measurements on ACL TOP displayed approximately double CV% compared to Stago sthemO, irrespective of hemolysis.CONCLUSIONS: Both analyzers showed comparable performance for PT and APTT in hemolyzed and non-hemolyzed samples, while fibrinogen displayed a different trend. Mechanical clot detection provided better overall precision for fibrinogen and for APTT in hemolyzed specimens.

Biochimica ClinicaVol. 50(1)
University of Verona (IT), Charles Sturt University (AU), Westmead Hospital (AU)
Openalex Percentile: Top 64%
Clinical Laboratory Practices and Quality Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.