What is the proportion of samples reported as positive among reactive samples in Siemens Atellica IM analyzer HIV and HCV serological screening?

• 69% of reactive HIV Ag/Ab samples with Atellica IM analyzer were reported positive • 96% of reactive anti-HCV samples with Atellica IM analyzer were reported positive • Reactive samples over measurement range were positive with high posterior probability Background: Serological screening of transmissible blood-borne viruses, like HIV and HCV, is a seminal part of medical laboratory workflow. Reliable screening methods are vital to provide impetus for confirmation testing, since false reactive results occur with HIV and HCV immunoassays. Objectives: We analyzed what proportion of the reactive HIV Ag/Ab and anti-HCV serum samples with Siemens Atellica IM Analyzer were finally reported as positive after confirmation testing. Study design: Reactive samples with Atellica IM Analyzers using HIV Ag/Ab Combo Assay (HIV Ag/Ab) or Atellica IM Hepatitis C assay (anti-HCV) analyzed between 1/2021-4/2023 were queried from our laboratory system. Index cut-off for reactive sample was ≥1.00 for HIV Ag/Ab and ≥0.80 for anti-HCV. All reactive samples underwent confirmation testing and were interpreted by clinical microbiologists. We calculated the proportion of positive results amidst reactive samples with Atellica IM Analyzers. We used index categorization based on our laboratory workflow for confirmation testing. Results: Of 802 reactive HIV Ag/Ab and 3100 reactive anti-HCV samples, 69% (95% CI 66-72%) and 96% (95% CI 95-97%) were eventually reported positive. Reactive samples with an index over measurement range were almost exclusively reported as positive with high posterior probabilities. Reactive HIV Ag/Ab samples within a very low or low reactivity category were seldom reported as positive. Conclusions: Most reactive samples with Atellica IM Analyzer were reported positive, nonetheless up to 31% of HIV Ag/Ab and 4% of anti-HCV samples were false reactives. These data can be utilized in designing algorithms for confirmation testing in clinical laboratories.

Authors

Institutions

Publication Details

Journal
Journal of Clinical Virology Plus
Published
2025-10-10
DOI
https://doi.org/10.1016/j.jcvp.2025.100232
Primary Topic
Probabilistic Statistics in Medicine
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

What is the proportion of samples reported as positive among reactive samples in Siemens Atellica IM analyzer HIV and HCV serological screening?

Jesper M. Kivelä, Anne Toivonen, Maarit J. Ahava, Satu Kurkela
Journal of Clinical Virology Plus
Probabilistic Statistics in Medicine
article

What is the proportion of samples reported as positive among reactive samples in Siemens Atellica IM analyzer HIV and HCV serological screening?

Jesper M. Kivelä, Anne Toivonen, Maarit J. Ahava, Satu Kurkela
article en

Abstract

• 69% of reactive HIV Ag/Ab samples with Atellica IM analyzer were reported positive • 96% of reactive anti-HCV samples with Atellica IM analyzer were reported positive • Reactive samples over measurement range were positive with high posterior probability Background: Serological screening of transmissible blood-borne viruses, like HIV and HCV, is a seminal part of medical laboratory workflow. Reliable screening methods are vital to provide impetus for confirmation testing, since false reactive results occur with HIV and HCV immunoassays. Objectives: We analyzed what proportion of the reactive HIV Ag/Ab and anti-HCV serum samples with Siemens Atellica IM Analyzer were finally reported as positive after confirmation testing. Study design: Reactive samples with Atellica IM Analyzers using HIV Ag/Ab Combo Assay (HIV Ag/Ab) or Atellica IM Hepatitis C assay (anti-HCV) analyzed between 1/2021-4/2023 were queried from our laboratory system. Index cut-off for reactive sample was ≥1.00 for HIV Ag/Ab and ≥0.80 for anti-HCV. All reactive samples underwent confirmation testing and were interpreted by clinical microbiologists. We calculated the proportion of positive results amidst reactive samples with Atellica IM Analyzers. We used index categorization based on our laboratory workflow for confirmation testing. Results: Of 802 reactive HIV Ag/Ab and 3100 reactive anti-HCV samples, 69% (95% CI 66-72%) and 96% (95% CI 95-97%) were eventually reported positive. Reactive samples with an index over measurement range were almost exclusively reported as positive with high posterior probabilities. Reactive HIV Ag/Ab samples within a very low or low reactivity category were seldom reported as positive. Conclusions: Most reactive samples with Atellica IM Analyzer were reported positive, nonetheless up to 31% of HIV Ag/Ab and 4% of anti-HCV samples were false reactives. These data can be utilized in designing algorithms for confirmation testing in clinical laboratories.

Journal of Clinical Virology PlusVol. 5(4)
University of Helsinki (FI), Helsinki University Hospital (FI), Hospital District of Helsinki and Uusimaa (FI)
Openalex Percentile: Top 73%
Probabilistic Statistics in Medicine
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.