Effective point of measurement calculations for cylindrical ionization chambers in kilovoltage x-ray beams

<i>Objective.</i>Knowledge of the effective point of measurement (EPOM) of ionization chambers is needed in relative dosimetry for kilovoltage x-ray beams. This study used Monte Carlo calculations to determine EPOM of three cylindrical ionization chambers in kilovoltage x-ray beams.<i>Approach.</i>The EGSnrc Monte Carlo code was used for all calculations for central axis depth doses in a water phantom and in the air cavity of cylindrical ionization chambers in water. EPOM shifts were calculated for x-ray beams from 50 kVp to 300 kVp with half value layer (HVL) from 0.07 to 3.5 mm Cu, with field sizes from 5 to 15 cm diameter. Three cylindrical ionization chambers (NE 2571, Exradin A12 and PTW 30013) were realistically modelled in EGSnrc. The EPOM shift value was identified from the minimum deviation in the ratio of the dose-to-water and the dose-to-air in the chamber cavity, following published numerical methods.<i>Main Results.</i>EPOM shifts normalized to the radius (r) of the cylindrical chamber of the air cavity were calculated to be within 0.2 r of the central axis for HVLs between 0.5 and 3.5 mm Cu. For HVLs less than 0.5 mm Cu, the EPOM shift increased with decreasing HVL. The EPOM shift did not change significantly with field size. Once the EPOM shift is applied, residual differences between the dose-to-water depth curves and dose-to-air cavity depth curves were found to be less than 0.5% in most cases.<i>Significance.</i>Monte Carlo calculations confirm a zero EPOM shift for cylindrical ionization chambers in x-ray beams with HVLs greater than 0.5 mm Cu. The calculated EPOM shifts begin to deviate significantly from zero for HVLs less than 0.5 mm Cu. Current recommendations of a uniform zero EPOM shift for cylindrical ionization chambers in kilovoltage x-ray beams may be inadequate for accurate relative dosimetry in the entire medium-energy x-ray range.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2025-10-09
DOI
https://doi.org/10.1088/1361-6560/ae1180
Primary Topic
Radiation Effects and Dosimetry
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Effective point of measurement calculations for cylindrical ionization chambers in kilovoltage x-ray beams

Robin Hill, Brendan Healy, Wei Tan
Physics in Medicine and Biology
Radiation Effects and Dosimetry
article

Effective point of measurement calculations for cylindrical ionization chambers in kilovoltage x-ray beams

Robin Hill, Brendan Healy, Wei Tan
article en

Abstract

<i>Objective.</i>Knowledge of the effective point of measurement (EPOM) of ionization chambers is needed in relative dosimetry for kilovoltage x-ray beams. This study used Monte Carlo calculations to determine EPOM of three cylindrical ionization chambers in kilovoltage x-ray beams.<i>Approach.</i>The EGSnrc Monte Carlo code was used for all calculations for central axis depth doses in a water phantom and in the air cavity of cylindrical ionization chambers in water. EPOM shifts were calculated for x-ray beams from 50 kVp to 300 kVp with half value layer (HVL) from 0.07 to 3.5 mm Cu, with field sizes from 5 to 15 cm diameter. Three cylindrical ionization chambers (NE 2571, Exradin A12 and PTW 30013) were realistically modelled in EGSnrc. The EPOM shift value was identified from the minimum deviation in the ratio of the dose-to-water and the dose-to-air in the chamber cavity, following published numerical methods.<i>Main Results.</i>EPOM shifts normalized to the radius (r) of the cylindrical chamber of the air cavity were calculated to be within 0.2 r of the central axis for HVLs between 0.5 and 3.5 mm Cu. For HVLs less than 0.5 mm Cu, the EPOM shift increased with decreasing HVL. The EPOM shift did not change significantly with field size. Once the EPOM shift is applied, residual differences between the dose-to-water depth curves and dose-to-air cavity depth curves were found to be less than 0.5% in most cases.<i>Significance.</i>Monte Carlo calculations confirm a zero EPOM shift for cylindrical ionization chambers in x-ray beams with HVLs greater than 0.5 mm Cu. The calculated EPOM shifts begin to deviate significantly from zero for HVLs less than 0.5 mm Cu. Current recommendations of a uniform zero EPOM shift for cylindrical ionization chambers in kilovoltage x-ray beams may be inadequate for accurate relative dosimetry in the entire medium-energy x-ray range.

Physics in Medicine and BiologyVol. 70(21)
The University of Sydney (AU), Gosford Hospital (AU), Environmental Protection Agency (DK), Australian Radiation Protection and Nuclear Safety Agency
Openalex Percentile: Top 66%
Radiation Effects and Dosimetry
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.