The unique properties and versatile applications of laser-induced graphene (LIG) have garnered significant interest for electrochemical sensing technologies. In this study, we report the fabrication and application of an in-house produced LIG/polyimide (PI) composite, generated via 450 nm laser irradiation, for the amperometric detection of Nicotinamide Adenine Dinucleotide (NADH), a critical biomarker associated with several neurodegenerative human diseases. The LIG structure was confirmed by Raman spectroscopy, X-ray photoelectron spectroscopy (XPS), and sheet resistance measurements, with an average sheet resistance (<i>R</i> <sub>s</sub>) of 24.38 ± 2.19 Ω/□, indicating excellent electrical conductivity. XPS analysis revealed the presence of CO bonds (288.9 eV), formed under oxidizing conditions during LIG fabrication, which may contribute to enhanced electrocatalytic activity by facilitating NADH oxidation through redox mediation. Using a printed Ag/AgCl pseudoreference electrode and an applied potential of only 50 mV vs Ag/AgCl, NADH was detected within a concentration range of 5 μmol L<sup>-1</sup> to 10 mmol L<sup>-1</sup>. The sensor exhibited a limit of detection (LOD) of 2.72 μmol L<sup>-1</sup> and a limit of quantification (LOQ) of 9.07 μmol L<sup>-1</sup>, with a linear response up to 1 mmol L<sup>-1</sup>. The repeatability of the electrochemical oxidation of NADH resulted in a relative standard deviation (RSD) of 2.76%, while the reproducibility, evaluated as intra- and interbatch variability, yielded RSD values of 5.78 and 8.22%, respectively. Furthermore, the total material cost for each electrochemical cell was estimated at only U$ 0.10, highlighting the method's potential for low-cost and environmentally friendly biosensor development. The fabricated LIG platform offers a promising route for sensitive, scalable, and sustainable detection of NADH and potentially other clinically relevant analytes.

Authors

Institutions

Publication Details

Journal
ACS Omega
Published
2025-10-09
DOI
https://doi.org/10.1021/acsomega.5c04282
Primary Topic
Electrochemical sensors and biosensors
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Cost-Effective Fabrication of Laser-Induced Graphene Electrochemical Cell for NADH Detection

Cecília de Carvalho Castro Silva, Camila M. Maroneze, Matheus Santos Dias, Tatiana Parra Vello et al.
ACS Omega
Electrochemical sensors and biosensors
article

Cost-Effective Fabrication of Laser-Induced Graphene Electrochemical Cell for NADH Detection

Cecília de Carvalho Castro Silva, Camila M. Maroneze, Matheus Santos Dias, Tatiana Parra Vello, Elsa M. Materón, Ketley Caroline Rocha Pereira, Gustavo Miguel Sousa, Deissy Feria Garnica
article en

Abstract

The unique properties and versatile applications of laser-induced graphene (LIG) have garnered significant interest for electrochemical sensing technologies. In this study, we report the fabrication and application of an in-house produced LIG/polyimide (PI) composite, generated via 450 nm laser irradiation, for the amperometric detection of Nicotinamide Adenine Dinucleotide (NADH), a critical biomarker associated with several neurodegenerative human diseases. The LIG structure was confirmed by Raman spectroscopy, X-ray photoelectron spectroscopy (XPS), and sheet resistance measurements, with an average sheet resistance (<i>R</i> <sub>s</sub>) of 24.38 ± 2.19 Ω/□, indicating excellent electrical conductivity. XPS analysis revealed the presence of CO bonds (288.9 eV), formed under oxidizing conditions during LIG fabrication, which may contribute to enhanced electrocatalytic activity by facilitating NADH oxidation through redox mediation. Using a printed Ag/AgCl pseudoreference electrode and an applied potential of only 50 mV vs Ag/AgCl, NADH was detected within a concentration range of 5 μmol L<sup>-1</sup> to 10 mmol L<sup>-1</sup>. The sensor exhibited a limit of detection (LOD) of 2.72 μmol L<sup>-1</sup> and a limit of quantification (LOQ) of 9.07 μmol L<sup>-1</sup>, with a linear response up to 1 mmol L<sup>-1</sup>. The repeatability of the electrochemical oxidation of NADH resulted in a relative standard deviation (RSD) of 2.76%, while the reproducibility, evaluated as intra- and interbatch variability, yielded RSD values of 5.78 and 8.22%, respectively. Furthermore, the total material cost for each electrochemical cell was estimated at only U$ 0.10, highlighting the method's potential for low-cost and environmentally friendly biosensor development. The fabricated LIG platform offers a promising route for sensitive, scalable, and sustainable detection of NADH and potentially other clinically relevant analytes.

ACS OmegaVol. 10(41)
Universidade Presbiteriana Mackenzie (BR)
Openalex Percentile: Top 72%
Electrochemical sensors and biosensors
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.