LA-ICP-TOF-MS for quantitative mapping of biogenic carbonate samples using matrix-matched nanoparticulate pressed powder pellets

This study evaluated the micro-homogeneity of seven different commercially available nanoparticulate pressed pellets based on a CaCO<sub>3</sub> matrix and their utility for quantitative elemental mapping of biogenic carbonates using laser ablation-inductively coupled plasma-time-of-flight-mass spectrometry (LA-ICP-TOF-MS). The analytical performance of matrix-matched calibration (using the aforementioned nano-pellets) was compared against that of non-matrix-matched calibration using a silicate glass (NIST SRM 610) reference material for quantification. Calibration with nano-pellets, combined with the use of Ca as an internal standard, significantly improved the quantification accuracy, providing recoveries between 80-120% for the majority of the 18 elements selected and spread across a wide concentration range (from sub-μg g<sup>-1</sup> to tens of wt%). However, some nano-pellets (<i>e.g.</i>, BPLM-NP and BAM-RS3-NP) exhibited higher heterogeneity, leading to biased recoveries. Also, an inverse correlation between the mass fraction and the relative standard deviation (RSD) was observed. Throughout the work, elemental mapping was conducted with a laser beam size of 20 × 20 μm<sup>2</sup> as a compromise between spatial resolution, sensitivity (sub-μg g<sup>-1</sup> limits of detection required for trace elements), linear dynamic range, total analysis time and size of the region-of-interest. The quantitative mapping approach enabled the generation of high-resolution, multi-elemental 2D-maps of various CaCO<sub>3</sub>-based natural chronological archives, including fish otoliths and bivalve shells, revealing detailed elemental distribution patterns for both trace (<i>e.g.</i>, Mn, Ba) and major elements (<i>e.g.</i>, Sr). This LA-ICP-TOF-MS methodology provides a powerful tool for resolving intricate microstructures and thus, for chronologically tracking bioaccumulation of environmentally relevant metals, offering significant advantages over traditional 1D (line scanning) analysis as the latter may lead to misinterpretation of elemental distributions.

Authors

Institutions

Publication Details

Journal
Journal of Analytical Atomic Spectrometry
Published
2025-01-01
DOI
https://doi.org/10.1039/d5ja00280j
Primary Topic
Calcium Carbonate Crystallization and Inhibition
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

LA-ICP-TOF-MS for quantitative mapping of biogenic carbonate samples using matrix-matched nanoparticulate pressed powder pellets

Frank Vanhaecke, Simon Nordstad, Martin Wiech, Ana Lores‐Padín et al.
Journal of Analytical Atomic Spectrometry
Calcium Carbonate Crystallization and Inhibition
article

LA-ICP-TOF-MS for quantitative mapping of biogenic carbonate samples using matrix-matched nanoparticulate pressed powder pellets

Frank Vanhaecke, Simon Nordstad, Martin Wiech, Ana Lores‐Padín, Thibaut Van Acker, Niels J. de Winter, Yannic Hallier
article en

Abstract

This study evaluated the micro-homogeneity of seven different commercially available nanoparticulate pressed pellets based on a CaCO<sub>3</sub> matrix and their utility for quantitative elemental mapping of biogenic carbonates using laser ablation-inductively coupled plasma-time-of-flight-mass spectrometry (LA-ICP-TOF-MS). The analytical performance of matrix-matched calibration (using the aforementioned nano-pellets) was compared against that of non-matrix-matched calibration using a silicate glass (NIST SRM 610) reference material for quantification. Calibration with nano-pellets, combined with the use of Ca as an internal standard, significantly improved the quantification accuracy, providing recoveries between 80-120% for the majority of the 18 elements selected and spread across a wide concentration range (from sub-μg g<sup>-1</sup> to tens of wt%). However, some nano-pellets (<i>e.g.</i>, BPLM-NP and BAM-RS3-NP) exhibited higher heterogeneity, leading to biased recoveries. Also, an inverse correlation between the mass fraction and the relative standard deviation (RSD) was observed. Throughout the work, elemental mapping was conducted with a laser beam size of 20 × 20 μm<sup>2</sup> as a compromise between spatial resolution, sensitivity (sub-μg g<sup>-1</sup> limits of detection required for trace elements), linear dynamic range, total analysis time and size of the region-of-interest. The quantitative mapping approach enabled the generation of high-resolution, multi-elemental 2D-maps of various CaCO<sub>3</sub>-based natural chronological archives, including fish otoliths and bivalve shells, revealing detailed elemental distribution patterns for both trace (<i>e.g.</i>, Mn, Ba) and major elements (<i>e.g.</i>, Sr). This LA-ICP-TOF-MS methodology provides a powerful tool for resolving intricate microstructures and thus, for chronologically tracking bioaccumulation of environmentally relevant metals, offering significant advantages over traditional 1D (line scanning) analysis as the latter may lead to misinterpretation of elemental distributions.

Journal of Analytical Atomic SpectrometryVol. 40(11)
Norwegian Institute of Marine Research (NO), Ghent University (BE), Clinical Research Center Kiel (DE), Vrije Universiteit Amsterdam (NL)
Fonds Wetenschappelijk Onderzoek, Nederlandse Organisatie voor Wetenschappelijk Onderzoek, Vlaamse regering
Openalex Percentile: Top 82%
Calcium Carbonate Crystallization and Inhibition
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.