Impact of Lower Limb Amputation on Implant Heating During 1. <scp>5 T MRI</scp> : Assessment of Active and Passive Device
Limb amputation significantly alters RF field distribution, affecting RF-induced heating near both active and passive implants. These results highlight the need for amputation-specific safety assessments during MRI to ensure accurate evaluation of heating risks and implant compatibility in amputee patients.
Authors
- Lingfei Zhang (ORCID: https://orcid.org/0000-0002-9831-8708)
- Jianfeng Zheng (ORCID: https://orcid.org/0000-0001-9981-8256)
- Mir Khadiza Akter
- Michael Steckner (ORCID: https://orcid.org/0000-0001-6854-9348)
- Ji Chen (ORCID: https://orcid.org/0000-0001-6944-2729)
- Ao Shen
- Mingtao Du (ORCID: https://orcid.org/0009-0002-1548-8785)
- Ziyu Zuo
Institutions
- University of Houston (US)
Publication Details
- Journal
- Magnetic Resonance in Medicine
- Published
- 2025-10-08
- DOI
- https://doi.org/10.1002/mrm.70123
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00