Abstract This study attempts to design and develop microdosimeters with optimal accuracy based on alternative gases to make an accurate microdosimeter. So we examined alternative gases, and all the effects of these gases on neutron microdosimetric spectra studied. In this research, using the Geant4-10-4 code we simulated the neutron point source with energies of 0.1, 1, and 10 MeV. To compare between different states, total microdosimetric quantities, and these statistical uncertainties in each case were calculated and reported. To replace gases using the SRIM (Stopping and Range of Ions in Matter) code data and assuming that the stopping power curve in terms of energy for methane and propane gases is similar to the methane and propane based tissue-equivalent gases, we calculated density correction factors. These correction factors provided suitable conditions for the replacement of the methane and propane gases instead of the methane and propane-based tissue equivalent gas in the microdosimetric cavity. The density correction factors for methane and propane gases are 0.781 and 0.864 respectively. Therefore, we were able to use these alternative gases to make this precise and practical device with good agreement and with a small difference in the microdosimetric spectra.

Authors

Institutions

Publication Details

Journal
Journal of Instrumentation
Published
2025-10-01
DOI
https://doi.org/10.1088/1748-0221/20/10/p10019
Primary Topic
Nuclear Physics and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Calculation of the correction factor of alternative gases on the microdosimetric spectra at different neutron energies

Somayeh Jahanfar, Hossein Tavakoli-Anbaran
Journal of Instrumentation
Nuclear Physics and Applications
article

Calculation of the correction factor of alternative gases on the microdosimetric spectra at different neutron energies

Somayeh Jahanfar, Hossein Tavakoli-Anbaran
article en

Abstract

Abstract This study attempts to design and develop microdosimeters with optimal accuracy based on alternative gases to make an accurate microdosimeter. So we examined alternative gases, and all the effects of these gases on neutron microdosimetric spectra studied. In this research, using the Geant4-10-4 code we simulated the neutron point source with energies of 0.1, 1, and 10 MeV. To compare between different states, total microdosimetric quantities, and these statistical uncertainties in each case were calculated and reported. To replace gases using the SRIM (Stopping and Range of Ions in Matter) code data and assuming that the stopping power curve in terms of energy for methane and propane gases is similar to the methane and propane based tissue-equivalent gases, we calculated density correction factors. These correction factors provided suitable conditions for the replacement of the methane and propane gases instead of the methane and propane-based tissue equivalent gas in the microdosimetric cavity. The density correction factors for methane and propane gases are 0.781 and 0.864 respectively. Therefore, we were able to use these alternative gases to make this precise and practical device with good agreement and with a small difference in the microdosimetric spectra.

Journal of InstrumentationVol. 20(10)
University of Neyshabur (IR), University of Shahrood (IR)
Openalex Percentile: Top 62%
Nuclear Physics and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.