The 3D NPS provides new insight into the noise characteristics of PCD-CT: Noise in the native PCD-CT projection data is uncorrelated across detector rows, but clinical reconstruction processes introduce noise spatial correlation along both axial and z directions, particularly at higher QIR strengths, lower radiation doses, or with larger image objects that increase the percentage of scattered photons. Compared to the Standard mode, UHR mode reconstruction exhibits stronger noise correlation due to its superior detector spatial resolution, allowing for more aggressive spatial smoothing to achieve the desired spatial resolution in the final image.

Authors

Institutions

Publication Details

Journal
Medical Physics
Published
2025-10-01
DOI
https://doi.org/10.1002/mp.70067
Citations
2
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
1.35
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Three‐dimensional noise characteristics of clinical photon counting detector CT

Corey T. Jensen, Megan C. Jacobsen, Frank Dong, Xinming Liu et al.
2 citations
Medical Physics
Advanced X-ray and CT Imaging
1.35
article

Three‐dimensional noise characteristics of clinical photon counting detector CT

Corey T. Jensen, Megan C. Jacobsen, Frank Dong, Xinming Liu, Humberto Monsivais, Ke Li, John Rong
article en
2 citations

Abstract

The 3D NPS provides new insight into the noise characteristics of PCD-CT: Noise in the native PCD-CT projection data is uncorrelated across detector rows, but clinical reconstruction processes introduce noise spatial correlation along both axial and z directions, particularly at higher QIR strengths, lower radiation doses, or with larger image objects that increase the percentage of scattered photons. Compared to the Standard mode, UHR mode reconstruction exhibits stronger noise correlation due to its superior detector spatial resolution, allowing for more aggressive spatial smoothing to achieve the desired spatial resolution in the final image.

Medical PhysicsVol. 52(10)
The University of Texas MD Anderson Cancer Center (US)
Openalex Percentile: Top 17%
Advanced X-ray and CT Imaging
1.35
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.