Progressive X-ray tube degradation detected via daily mammography quality control

This study evaluated whether routine daily quality control (QC) tests can anticipate X-ray tube degradation in a digital mammography system. Over 30 months, tube loading, contrast-to-noise ratio (CNR), and image quality scores were extracted from daily QC tests and analysed alongside a documented tube failure in September 2024. These results were compared with annual tube output measurements and environmental data. A progressive rise in tube loading was observed prior to tube failure, increasing from 84.45 to 94.58 mAs (12.0%). The Mann-Kendall test confirmed a significant upward trend (τ = 0.461; <i>p</i> < 0.001), with Sen's slope indicating +0.19 mAs/month. Despite this, CNR values remained stable (mean = 14.76 ± 0.20), and image quality showed no degradation. However, tube output declined consistently (from 30.95 ± 3.08 to 28.76 ± 0.36 μGy/mAs), confirming reduced efficiency. Temperature and humidity remained within recommended limits. These findings suggest that gradual increases in tube loading can indicate early tube degradation, even when image quality remains unaffected. Beyond confirming equipment performance, daily QC may serve as a strategic tool to detect early signs of degradation, extend device lifespan, reduce unplanned downtime, and promote patient safety without added costs.

Authors

Institutions

Publication Details

Journal
Journal of Medical Engineering & Technology
Published
2025-10-07
DOI
https://doi.org/10.1080/03091902.2025.2553132
Primary Topic
Digital Radiography and Breast Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Progressive X-ray tube degradation detected via daily mammography quality control

Túlio Guilherme Soares Marques, Matheus Alvarez, Diana Rodrigues de Pina
Journal of Medical Engineering & Technology
Digital Radiography and Breast Imaging
article

Progressive X-ray tube degradation detected via daily mammography quality control

Túlio Guilherme Soares Marques, Matheus Alvarez, Diana Rodrigues de Pina
article en

Abstract

This study evaluated whether routine daily quality control (QC) tests can anticipate X-ray tube degradation in a digital mammography system. Over 30 months, tube loading, contrast-to-noise ratio (CNR), and image quality scores were extracted from daily QC tests and analysed alongside a documented tube failure in September 2024. These results were compared with annual tube output measurements and environmental data. A progressive rise in tube loading was observed prior to tube failure, increasing from 84.45 to 94.58 mAs (12.0%). The Mann-Kendall test confirmed a significant upward trend (τ = 0.461; <i>p</i> < 0.001), with Sen's slope indicating +0.19 mAs/month. Despite this, CNR values remained stable (mean = 14.76 ± 0.20), and image quality showed no degradation. However, tube output declined consistently (from 30.95 ± 3.08 to 28.76 ± 0.36 μGy/mAs), confirming reduced efficiency. Temperature and humidity remained within recommended limits. These findings suggest that gradual increases in tube loading can indicate early tube degradation, even when image quality remains unaffected. Beyond confirming equipment performance, daily QC may serve as a strategic tool to detect early signs of degradation, extend device lifespan, reduce unplanned downtime, and promote patient safety without added costs.

Journal of Medical Engineering & TechnologyVol. 50(1)
Universidade Estadual Paulista (Unesp) (BR)
Openalex Percentile: Top 59%
Digital Radiography and Breast Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Progressive X-ray tube degradation detected via daily mammography quality control — Túlio Guilherme Soares Marques, Matheus Alvarez, et al. · Journal of Medical Engineering & Technology (2025) | TGRS Research Map | TGRS