Validation of the automated oscillometric upper-arm cuff home blood pressure monitor Microlife BP3KT1-4X (BP B3 Comfort PC) in a general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypert
The automated oscillometric home BP monitor Microlife BP3KT1-4X (BP B3 Comfort PC) fulfilled all the requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018) and its amendments (1:2020 and 2:2024) in a general population and therefore can be recommended for self-monitoring of BP at home.
Authors
- Vasileios Ntousopoulos
- George S. Stergiou (ORCID: https://orcid.org/0000-0002-6132-0038)
- Konstantinos G. Kyriakoulis (ORCID: https://orcid.org/0000-0001-8986-2704)
- Ariadni Menti (ORCID: https://orcid.org/0000-0003-4258-9928)
- Αναστάσιος Κόλλιας (ORCID: https://orcid.org/0000-0002-2098-1665)
- Aikaterini Komnianou
- Dimitrios Mariglis
- Petroula Lempidaki
- Konstantina-Rozalia Bakalakou
Institutions
- National and Kapodistrian University of Athens (GR)
- Sotiria General Hospital (GR)
Publication Details
- Journal
- Blood Pressure Monitoring
- Published
- 2025-09-26
- DOI
- https://doi.org/10.1097/mbp.0000000000000773
- Primary Topic
- Blood Pressure and Hypertension Studies
- Type
- article
- Field-Weighted Citation Impact
- 0.00