A falsely low lactate: discrepancies in lactate measurement between blood gas and core laboratory instruments due to β-hydroxybutyrate interference—a case report
Dilution and spike experiments suggested potential interferences in core laboratory instruments using the Trinder reaction, leading to inaccurate lactate readings.