Sensor-augmented pumps (SAPs) and automated insulin delivery (AID) systems are innovative technologies for diabetes management. Accurate continuous glucose monitoring (CGM) is crucial for their safe and effective use; however, certain commonly used drugs can interfere with CGM accuracy. Although acetaminophen is known to cause falsely elevated CGM glucose values, previous CGM studies have primarily focused on its oral administration, with limited data on intravenous use. We report a case of a CGM reaction after the intravenous administration of acetaminophen in a boy with type 1 diabetes using SAP. The patient received repeated doses of intravenous acetaminophen (15 mg/kg for 15 min) for pain relief. After administration, we recorded a rapid increase in his CGM readings without a corresponding increase in blood glucose levels. The CGM glucose peaked at 29.2 ± 1.9 min (mean ± standard deviation) after administration and an estimated discrepancy of 55 to 114 mg/dL compared with capillary blood glucose measurements. Discrepancies between measured blood glucose and CGM readings were significantly greater at lower glucose levels. These falsely elevated CGM readings could potentially trigger an inappropriate autocorrection bolus in AID systems and increase the risk of hypoglycemia. Medical professionals should be fully aware of acetaminophen-induced CGM interference, particularly the potential risks in patients using AID systems.

Authors

Institutions

Publication Details

Journal
JMA Journal
Published
2025-01-01
DOI
https://doi.org/10.31662/jmaj.2025-0186
Primary Topic
Electrochemical sensors and biosensors
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Interference of Intravenous Acetaminophen with Continuous Glucose Monitoring System

Ryuta Masuya, Satoru Meiri, Kazuhiko Nakame, Misayo Matsuyama et al.
JMA Journal
Electrochemical sensors and biosensors
article

Interference of Intravenous Acetaminophen with Continuous Glucose Monitoring System

Ryuta Masuya, Satoru Meiri, Kazuhiko Nakame, Misayo Matsuyama, Hiroshi Moritake, Hirotake Sawada
article en

Abstract

Sensor-augmented pumps (SAPs) and automated insulin delivery (AID) systems are innovative technologies for diabetes management. Accurate continuous glucose monitoring (CGM) is crucial for their safe and effective use; however, certain commonly used drugs can interfere with CGM accuracy. Although acetaminophen is known to cause falsely elevated CGM glucose values, previous CGM studies have primarily focused on its oral administration, with limited data on intravenous use. We report a case of a CGM reaction after the intravenous administration of acetaminophen in a boy with type 1 diabetes using SAP. The patient received repeated doses of intravenous acetaminophen (15 mg/kg for 15 min) for pain relief. After administration, we recorded a rapid increase in his CGM readings without a corresponding increase in blood glucose levels. The CGM glucose peaked at 29.2 ± 1.9 min (mean ± standard deviation) after administration and an estimated discrepancy of 55 to 114 mg/dL compared with capillary blood glucose measurements. Discrepancies between measured blood glucose and CGM readings were significantly greater at lower glucose levels. These falsely elevated CGM readings could potentially trigger an inappropriate autocorrection bolus in AID systems and increase the risk of hypoglycemia. Medical professionals should be fully aware of acetaminophen-induced CGM interference, particularly the potential risks in patients using AID systems.

JMA JournalVol. 8(4)
University of Miyazaki (JP), University of Miyazaki Hospital (JP)
Openalex Percentile: Top 73%
Electrochemical sensors and biosensors
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.