Detection of Typical and Blast Cells in Acute and Chronic Leukaemia Using the Manual Method and Hematology Analyzer BC-720
Introduction: Diagnosing Leukaemia requires precise identification of atypical whiteblood cells. This research aims to compare the effectiveness of the BC-720 analyzerand manual microscopy in identifying blast and typical cells in both acute and chronicLeukaemia. Manual microscopy and the BC-720 analyzer were used to investigatedifferential white blood cell counts from 68 Leukaemia patients and 18 controls.Sensitivity and error rates for detecting abnormal cells were evaluated. The manualapproach had a recognition loss of 12%, but successfully identified 47.1% of cases. TheBC-720 identified 52.9% of cases, with higher detection rates in chronic patients(41.2%) and control cases (5.9%). In acute cases, the detection loss was 8.0% lower.Both methods showed advantages and limitations. The BC-720 demonstrated higherdetection rates for abnormal cells, particularly in acute Leukaemia, and lowerrecognition losses compared to manual microscopy. It can aid in more accurateLeukaemia classification and management.
Authors
- Maher Matar
- Mehmed M. Bakhtaran
- Malak M. Keiba
Institutions
- Misurata University (LY)
- Libyan Academy (LY)
Publication Details
- Journal
- Misurata Medical Sciences Journal
- Published
- 2025-06-24
- DOI
- https://doi.org/10.36602/mmsj/2023.n09.06
- Primary Topic
- Digital Imaging for Blood Diseases
- Type
- article
- Field-Weighted Citation Impact
- 0.00