EFFECTIVENESS EVALUATION OF AUTOCLAVES AT THE NATIONAL INSTITUTE FOR CONTROL VACCINE AND BIOLOGICALS

National Institute for Control of Vaccines and Biologicals (NICVB) has utilized wet-heat sterilization methods for laboratory instruments, solution, and culture media. According to World Health Organization (WHO) recommendations, autoclaves used in sterilization processes require annual revalidation. Therefore, we conducted the study with the titled “Effectiveness Evaluation of Autoclaves at the National Institute for Control of Vaccines and Biologicals” to assess the operational performance and sterilization efficacy of autoclaves. The study was performed on two autoclaves: AC 10-TN and AC 11-TN. The results indicated that both autoclaves met operational performance requirements based on physical indicators, including: pressure ≥ 0.1 MPa, temperature ≥ 121°C, holding time ≥ 15 minutes and uniform heat distribution. Sterilization efficacy was also evaluated through biological indicators such as no appearance of microorganisms in both BI cultures and media that ensured sterility, while all positive controls met the required criteria and the risk of false negatives was eliminated. Consequently, both autoclaves were confirmed to be in high quality, reliable, and suitable for use in vaccine and biological product quality control testing

Authors

Publication Details

Journal
JOURNAL OF CONTROL VACCINES AND BIOLOGICALS
Published
2025-10-03
DOI
https://doi.org/10.56086/jcvb.v5i3.227
Primary Topic
Vaccine Coverage and Hesitancy
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

EFFECTIVENESS EVALUATION OF AUTOCLAVES AT THE NATIONAL INSTITUTE FOR CONTROL VACCINE AND BIOLOGICALS

Nguyễn Thị Quỳnh, Lê Thanh Thảo, Nguyen Ky Anh, Phạm Thị Hằng
JOURNAL OF CONTROL VACCINES AND BIOLOGICALS
Vaccine Coverage and Hesitancy
article

EFFECTIVENESS EVALUATION OF AUTOCLAVES AT THE NATIONAL INSTITUTE FOR CONTROL VACCINE AND BIOLOGICALS

Nguyễn Thị Quỳnh, Lê Thanh Thảo, Nguyen Ky Anh, Phạm Thị Hằng
article en

Abstract

National Institute for Control of Vaccines and Biologicals (NICVB) has utilized wet-heat sterilization methods for laboratory instruments, solution, and culture media. According to World Health Organization (WHO) recommendations, autoclaves used in sterilization processes require annual revalidation. Therefore, we conducted the study with the titled “Effectiveness Evaluation of Autoclaves at the National Institute for Control of Vaccines and Biologicals” to assess the operational performance and sterilization efficacy of autoclaves. The study was performed on two autoclaves: AC 10-TN and AC 11-TN. The results indicated that both autoclaves met operational performance requirements based on physical indicators, including: pressure ≥ 0.1 MPa, temperature ≥ 121°C, holding time ≥ 15 minutes and uniform heat distribution. Sterilization efficacy was also evaluated through biological indicators such as no appearance of microorganisms in both BI cultures and media that ensured sterility, while all positive controls met the required criteria and the risk of false negatives was eliminated. Consequently, both autoclaves were confirmed to be in high quality, reliable, and suitable for use in vaccine and biological product quality control testing

JOURNAL OF CONTROL VACCINES AND BIOLOGICALSVol. 5(3)
Openalex Percentile: Top 65%
Vaccine Coverage and Hesitancy
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.