CHARACTERIZATION OF AIRBORNE BACTERIAL COMMUNITIES AND EVALUATION OF THE DISINFECTION EFFICACY OF 5% HYDROGEN PEROXIDE IN ANIMAL HOUSING ROOMS AT THE LABORATORY ANIMAL DEPARTMENT

The composition of microorganisms in the air directly affects the quality of laboratory animals, staff, and determines the cleanliness level of the animal facility as well as the value of experiments. In order to eliminate pathogenic factors, 5% hydrogen peroxide was used at the Experimental Animals Department to disinfect animal housing rooms and auxiliary areas. The study results showed that the identified bacterial composition was mainly non-pathogenic Gram-positive bacteria under normal conditions, with staphylococci accounting for 72–92%, and other bacteria including Kocuria rhizophila, Enterococcus gallinarum, Bacillus atrophaeus, and Bacillus pumilus accounting for 8–28%. The average bacterial density in 12 sampled rooms mostly ranged was generally between 100–500 CFU/m³ of air, with no room exceeding 2000 CFU/m³. After disinfection, bacterial loads decreased by 4 to 30 times. Four weeks post-disinfection, all rooms maintained concentrations below 500 CFU/m³.

Authors

Publication Details

Journal
JOURNAL OF CONTROL VACCINES AND BIOLOGICALS
Published
2025-10-03
DOI
https://doi.org/10.56086/jcvb.v5i3.226
Primary Topic
Infection Control and Ventilation
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

CHARACTERIZATION OF AIRBORNE BACTERIAL COMMUNITIES AND EVALUATION OF THE DISINFECTION EFFICACY OF 5% HYDROGEN PEROXIDE IN ANIMAL HOUSING ROOMS AT THE LABORATORY ANIMAL DEPARTMENT

Nguyen Chi Hieu, Mẫn Thị Thành, Quách Thu Thảo, Trần Thị Hương Thơm et al.
JOURNAL OF CONTROL VACCINES AND BIOLOGICALS
Infection Control and Ventilation
article

CHARACTERIZATION OF AIRBORNE BACTERIAL COMMUNITIES AND EVALUATION OF THE DISINFECTION EFFICACY OF 5% HYDROGEN PEROXIDE IN ANIMAL HOUSING ROOMS AT THE LABORATORY ANIMAL DEPARTMENT

Nguyen Chi Hieu, Mẫn Thị Thành, Quách Thu Thảo, Trần Thị Hương Thơm, Tran Thi Hong, Hoang Trung Hung, Nguyễn Thị Thanh Nga, Dang Thi Luyen
article en

Abstract

The composition of microorganisms in the air directly affects the quality of laboratory animals, staff, and determines the cleanliness level of the animal facility as well as the value of experiments. In order to eliminate pathogenic factors, 5% hydrogen peroxide was used at the Experimental Animals Department to disinfect animal housing rooms and auxiliary areas. The study results showed that the identified bacterial composition was mainly non-pathogenic Gram-positive bacteria under normal conditions, with staphylococci accounting for 72–92%, and other bacteria including Kocuria rhizophila, Enterococcus gallinarum, Bacillus atrophaeus, and Bacillus pumilus accounting for 8–28%. The average bacterial density in 12 sampled rooms mostly ranged was generally between 100–500 CFU/m³ of air, with no room exceeding 2000 CFU/m³. After disinfection, bacterial loads decreased by 4 to 30 times. Four weeks post-disinfection, all rooms maintained concentrations below 500 CFU/m³.

JOURNAL OF CONTROL VACCINES AND BIOLOGICALSVol. 5(3)
Openalex Percentile: Top 60%
Infection Control and Ventilation
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

CHARACTERIZATION OF AIRBORNE BACTERIAL COMMUNITIES AND EVALUATION OF THE DISINFECTION EFFICACY OF 5% HYDROGEN PEROXIDE IN ANIMAL HOUSING ROOMS AT THE LABORATORY ANIMAL DEPARTMENT — Nguyen Chi Hieu, Mẫn Thị Thành, et al. · JOURNAL OF CONTROL VACCINES AND BIOLOGICALS (2025) | TGRS Research Map | TGRS